GENERAL THEORY OF OPERATION 

 

The 150-Watt Dual Halogen Illuminator general operation is as follows.  Please 
consult Fig. 1 for wiring information. 

 
 
 
 

A.C. POWER DISTRIBUTION 

 

The  150-Watt  Dual  Halogen  Illuminators  are  based  around  a  switch  able  input 
power transformer.  Input line voltages of 100-120 V ˜ and 220-240V  ˜ at 50/60 
Hz are applied at the Power Input Model via a Hospital grade power cord.  The 
voltage selector switch located on the rear panel selects the input power.  Over 
current  protection  is  provided  by  two  5-amp  circuit  breakers.    A  double-pole 
interlock  switch  provides  operator  safety,  which  is  located  on  right  side  of  the 
lamp heat shield.   
 
 

INTENSITY CONTROL  

 

A rotating stainless steel disc that is placed in front of the lamp controls intensity.  
The  disc  contains  holes  in  varying  sizes  and  patterns.    Manual  lamp  intensity 
control is made via a front panel mounted knob.   
 
 
 

COOLING SYSTEM 

 

Cooling is provided by a 76 CFM, 115-VAC fan (Refer to Internal View page five).  
Air is drawn through the sides of the unit, across the transformer and lamp, and 
exhausted  through  the  rear  mounted  exhaust  louver.  Caution  should  still  be 
used, as there are still potentially hazardous temperatures at the turret. 
 

 
 
 
 
 
 
 
 
 
 
 

LIT054CUDA 

 

Rev. A 

(English)  

 

Page 4 of 9  

Содержание MD-150T

Страница 1: ...LIT054CUDA Rev A English MD 150T 150 Watt Dual Halogen Illuminator Service Manual 6018 Bowdendale Avenue Jacksonville FL 32116 Customer Service 904 737 7611 FAX 904 733 4832 Toll Free 877 677 2832...

Страница 2: ...ICES 3 GENERAL THEORY OF OPERATION 4 INTERNAL VIEW 5 CIRCUIT BREAKER REPLACEMENT 6 SHUTTER REPLACEMENT 6 TRANSFORMER REPLACEMENT 7 COOLING FAN REPLACEMENT 7 ELECTRICAL SCHEMATIC Figure 1 8 REPLACEMENT...

Страница 3: ...circuitry and wiring which could be hazardous with contact Use extreme caution when working on equipment that has power applied to it TECHNICAL SUPPORT SERVICES In the event that you experience diffi...

Страница 4: ...circuit breakers A double pole interlock switch provides operator safety which is located on right side of the lamp heat shield INTENSITY CONTROL A rotating stainless steel disc that is placed in fro...

Страница 5: ...LIT054CUDA Rev A English Page 5 of 9...

Страница 6: ...1 Disconnect the illuminator from the power source 2 Remove the 12 cover screws in the sides of the unit 3 Remove the 3 screws on the bottom of the unit which holds the front panel to the bottom hous...

Страница 7: ...sure the lamp access door is over the lamp cartridge 11 Replace the 12 cover screws and secure COOLING FAN REPLACEMENT 1 Disconnect the illuminator from the power source 2 Remove the 12 cover screws i...

Страница 8: ...FIGURE 1 ELECTRICAL SCHEMATIC MD 150T DUAL HALOGEN ILLUMINATOR LIT054CUDA Rev A English Page 8 of 9...

Страница 9: ...5 175 20 800 018 Circuit Breaker 5A 800 015 Shutter Assembly 800 003 Cooling Fan SP103A 800 019 Shutter belt Large 800 020 Shutter belt Small 800 021 For parts not listed here call our Technical Servi...

Отзывы: