ADJUSTING THE SHIPPING BOLTS 

PLACEMENT AND LEVELING

1      Carefully tip the Cubiscan 110-T onto the back panel to expose the bottom of the scale.

2      Use a 13 mm open-ended wrench to turn all three of the shipping bolts clockwise until 
        the head is tight against the scale plate.

2      Place the bubble level provided in the center of the scale platter.
3      Adjust the five leveling feet under the Cubiscan 110-T to level the scale. Make sure all 
        feet are touching the supporting surface.

3      Tip the Cubiscan 110-T back to its upright position.

3      Use the screws from step 1 to secure the back panel in its upright position.

4      Insert the plug on each length sensor cable into its socket on the back (at each 
        end) of the Cubiscan 110-T.

5      Route the height sensor cable up the channel on the back panel, and insert the 
        plug into the socket on the height sensor.

1      Place the Cubiscan 110-T on a solid, stable surface. Be sure that nothing touches the 
        back wall or top platter, as this may affect the scale.