7. Operation 

Settings

Logger

Files

Print

Bluetooth 

Calibration

Service 

7.5 Data logger

Description of the function:

Start / Stop

Start and stop of the logger.

Sample rate

 

Set the time interval the samples will be recorded. For 
example 10 sec: The logger will record the measured 
values every 10 seconds.

Average

For long sampling rates, the user can choose to record 
the average value instead of the present value.

Logger status Status information about the logger.
Tester name User input for additional information about the person 

performing the measurement.

Location 

A location description can either be selected from a list or 
entered.

Company 
name 

Company name and address that should appear on print 
(usually the customer).

S 505

17

Содержание S 505

Страница 1: ...English Instruction and operation manual S 505 Dew point meter portable ...

Страница 2: ...e with this manual Should the device be tampered with in any manner other than a procedure which is described and specified in the manual the warranty is cancelled and the manufacturer is exempt from liability The device is destined exclusively for the described application CS iTEC offers no guarantee for the suitability for any other purpose CS iTEC is also not liable for consequential damage res...

Страница 3: ... 14 7 2 Symbols on screen 15 7 3 Main screen 16 7 4 Basic operation 16 7 5 Data logger 17 7 6 File menu 18 7 7 Print menu 18 7 8 Bluetooth menu 18 8 Application software 19 8 1 S4M S 19 8 2 S4C Handheld 19 9 Optional extra accessories 20 9 1 Portable bluetooth printer HDT 312 20 9 2 Parking measuring chamber by pass type 21 9 3 SD card 4G 22 10 Calibration 22 10 1 Dew point calibration 22 10 2 Pre...

Страница 4: ...ding this manual or the product please contact the manufacturer WARNING Compressed air Any contact with quickly escaping air or bursting parts of the compressed air system can lead to serious injuries or even death Only use pressure tight installation material Avoid that persons get hit escaping air or bursting parts of the instrument The system must be pressureless during maintenance work WARNING...

Страница 5: ...product in explosive areas Please observe the national regulations before during installation and operation Remarks It is not allowed to disassemble the product Always use spanner to mount the product properly ATTENTION Measurement values can be affected by malfunction The product must be installed properly and frequently maintained otherwise it may lead to wrong measurement values which can lead ...

Страница 6: ...1 Safety instructions For the storage the humidity has to be 90 no condensation 6 S 505 ...

Страница 7: ...ar The S 505 dew point meter is not developed to be used in explosive areas For the use in explosive areas please contact the manufacturer The S 505 dew point meter is mainly used in compressed air systems in industrial environment 3 Features Measures dew point temperature and pressure 3 sensor solution available Sensor A 100 Ctd 30 Ctd for trace moisture applications Sensor B 50 Ctd 50 Ctd for st...

Страница 8: ...easuring range Sensor A 100 Ctd 30 Ctd Sensor B 50 Ctd 50 Ctd Pressure 0 1 1 5 MPa Temperature 30 C 50 C Operating temperature 0 C 50 C Operating pressure 0 1 1 6 MPa Humidity of the meas medium 90 no condensation Casing material PC ABS Charging time 7 h if switched on 4 h if switched off Protection class IP65 Dimensions See dimensional drawing on the next page Display 2 8 colour graphic display W...

Страница 9: ...Data 4 3 Accuracy Accuracy Dew point 2 Ctd 50 C Temperature 0 3 C Pressure 0 005 MPa Repeatability 0 5 C Stated accuracy at Ambient process temperature of 23 C 3 C and ambient humidity of 90 no condensation S 505 9 ...

Страница 10: ...awing 5 Dimensional drawing 6 Installation Please make sure that all components listed below are included in your package Qty Description Item No 1 Handheld meter with data logger and S4M S software P560 0505 10 S 505 ...

Страница 11: ...tion certificate No P N 1 Instruction manual No P N 6 1 Installation Procedure The following steps explain the procedure of an appropriate installation 1 Define the proper sensor module which is used for the measurement This depends on the expected measurement value Use the sensor module A for expected dew point below 50 Ctd Use the sensor module B for expected dew point above 50 Ctd 2 To change t...

Страница 12: ... module for storage Remark The parking measuring chamber is mounted permanently on the sensor head to keep the sensor try For measuring the handle on the chamber has to be turned into measuring position After finishing the measurement please return to the park position Measuring position Parking position 12 S 505 ...

Страница 13: ...Connect the teflon hose with the measuring chamber The hose is connected to the 6 mm connector at the chamber 5 Connect the tip of the teflon hose with a quick connector The teflon hose with quick connector is used to connect the measuring chamber to the process Remark Please ensure that the teflon hose remains dry and clean Wet and contaminated hoses can effect the measurement and should be S 505...

Страница 14: ...n exchanged 6 2 Electrical connection Please recharge the dew point meter if the battery is empty For this use the charger which is included in your delivered package 7 Operation 7 1 Buttons and connectors 14 S 505 ...

Страница 15: ... of available energy The calibration is experted It is recommended to re calibrate the sensor unit For this pleas contact the manufacturer General error indication Please note down the error code and contact the manufacturer Bluetooth is active and ready to connect S 505 15 ...

Страница 16: ...gger is active The SD card is inserted 7 3 Main screen 7 4 Basic operation The operation is easy and similar Please try the functions through the touch screen and experience the features of this production On the left side of the screen the menu is shown It includes the follow menu items 16 S 505 ...

Страница 17: ...he measured values every 10 seconds Average For long sampling rates the user can choose to record the average value instead of the present value Logger status Status information about the logger Tester name User input for additional information about the person performing the measurement Location A location description can either be selected from a list or entered Company name Company name and add...

Страница 18: ...screen has to be used 7 6 File menu The file menu is used to view all recorded data We distinguish between two data files Single record data file can be printed with HDT 312 Multiple record data file can be analysed with CSM s 7 7 Print menu The print menu looks very much the same as the file menu but it will list only single recording files Only this files can be printed with the wireless printer...

Страница 19: ...rom the web page of the manufacturer 8 1 S4M S After download and installation the S 505 connect via USB to the computer Do following settings in S4M S Select Online to show current measurement values Select Download Data to download recorded data from S 505 to PC Select File List to open a data file for analysing Follow the instructions in Help File for more details Remark The S4M S requires a re...

Страница 20: ...al S 505 as pdf 9 Optional extra accessories 9 1 Portable bluetooth printer HDT 312 With the optional bluetooth printer it is possible to create a printout showing the measured values location and date time It is also possible to order the paper roll for the printer additional contains 3 rolls Principle Thermal Charging time 2 5 h Printing speed 60 mm s Stand by time 120 h Density 203 dpi Charger ...

Страница 21: ...est Select it for self test BT clear binding Select it to break the pairing Auto off time 30 m Language English Chinese 9 2 Parking measuring chamber by pass type This chamber can be used in applications where the measured gas is by passed through the chamber For more information please contact your retailer or the manufacturer S 505 21 ...

Страница 22: ...n the attached calibration certificate If the instrument has a slight drift due to ageing temperature or other effects a dew point and pressure calibration can be performed 10 1 Dew point calibration Performing a dew point calibration is critical and the following considerations should be taken Perform dew point calibration at the working point Foe example if you measure around 40 Ctd do the calib...

Страница 23: ... there is any doubt about correct calibration process and result the calibration can be removes at any time Please follow the instruction on the screen 10 2 Pressure calibration If the instrument is not showing 0 pressure at ambient condition it is recommended to perform a fast zero calibration For that purpose please use the calibration function and select pressure calibration Follow the steps on...

Страница 24: ...pose can also be carried by the manufacturer of the product for this please contact the manufacturer 13 Warranty CS iTEC provides a warranty for this product of 24 months covering the material and workmanship under the stated operating conditions from the date of delivery Please report any findings immediately and within the warranty time If faults occurring during the warranty time CS iTEC will r...

Страница 25: ...nstrument without a direct request written in this instruction manual If repairs or modifications are undertaken by third parties or unauthorised persons If the serial number has been changed damaged or removed Other claims especially those for damage occurring outside the instrument are not included unless responsibility is legally binding Warranty repairs do not extend the period of warranty ATT...

Страница 26: ...26 S 505 ...

Страница 27: ...S 505 27 ...

Страница 28: ... Sheung Shui N T Germany Hong Kong Tel 49 0 7631 936889 0 Tel 86 0 755 8619 3164 Fax 49 0 7631 936889 19 Fax 86 0 755 8619 3165 Email sales suto itec com Email sales cs itec asia Website http www suto itec com Website http www cs itec com All rights reserved Modifications and errors reserved S505_im_en_2017 2 ...