INSTALLATION
Please note: CS recommends the indirect installation with measuring chamber
Advantage
: Easy mounting and dismounting
of the probe without interruption of
the line. Quick response time due
to quick coupling. Optimum sensor
protection.
Indirectly in the compressed air system
Connect probe with measuring chamber to the
compessed air pipe by means of a quick
coupling. In case of compressed air containing
oil and dirt particles a pre-filter should be
installed in front of the measuring chamber.
Compressed air flows continuously (at 7 bar
approx. 1 l/min expanded) in the capillary pipe of
the measuring chamber. The reaction times for
the humidity reading are shorter than in case of
a direct mounting.
Directly in the compressed air system
Screw in probe with G 1/2" thread pressure-tight
in the center or at the top of the compressed air
pipe. Take care that measurement is effected
close to the compressed air flow. U-bend pipes
or non-flowing compressed air result in very
slow reaction times for the moisture reading.
Measurable gases
In general, humidity can be measured in all non-
corrosive gases. In case of measurements in
corrosive gases please consult CS Instruments
GmbH.