INSTALLATION 

 
 

Please note: CS recommends the indirect installation with measuring chamber 

 
 

Advantage

: Easy mounting and dismounting  

 

of the probe without interruption of 

 

the line. Quick response time due  

 

to quick coupling. Optimum sensor  

 protection. 
 
 
 
 

Indirectly in the compressed air system 

 
Connect probe with measuring chamber to the 
compessed air pipe by means of a quick 
coupling. In case of compressed air containing 
oil and dirt particles a pre-filter should be 
installed in front of the measuring chamber. 
Compressed air flows continuously (at 7 bar 
approx. 1 l/min expanded) in the capillary pipe of 
the measuring chamber. The reaction times for 
the humidity reading are shorter than in case of 
a direct mounting.  
 
 
 
 
 

Directly in the compressed air system 

 
Screw in probe with G 1/2" thread pressure-tight 
in the center or at the top of the compressed air 
pipe. Take care that measurement is effected 
close to the compressed air flow. U-bend pipes 
or non-flowing compressed air result in very 
slow reaction times for the moisture reading.  
 
 
 
 
 

Measurable gases 

 
In general, humidity can be measured in all non-
corrosive gases. In case of measurements in 
corrosive gases please consult CS Instruments 
GmbH. 
 

FA 410 

FA 410 

Содержание FA 410

Страница 1: ...A 4 41 10 0 Humidity measuring instruments for measurement of the pressure dew point and the atmospheric dew point in different applications compressed air plants refrigerating adsorption dryers gran...

Страница 2: ...ustomers buy our high standard products every year There are a few good reasons for doing so Cost performance ratio Reliable quality at a fair price We have the ideal solutions for your measuring task...

Страница 3: ...h accuracy of 2 Ctd Calibration on location and testing with CS control and calibration set PC connection set Programming via SFA Software Analogue output 4 20 mA scalable Calibration and adjustment i...

Страница 4: ...ard Power supply 16 30 VDC Output 4 20 mA 3 wire technology Protection class IP 65 EMV DIN EN 61326 Operating temperature 30 70 C ideal 0 50 C Storage temperature 40 80 C Load for analogue output 500...

Страница 5: ...onnection diagram DIAGRAM OF INSTRUMENT Pin 1 Pin 2 Pin 3 Pin 4 Pin 5 FA 410 Connector plug SDI VB VB I NC Connection cable 0554 0104 5 m 0554 0105 10 m brown white blue black SDI Digital signal inter...

Страница 6: ...uring chamber Compressed air flows continuously at 7 bar approx 1 l min expanded in the capillary pipe of the measuring chamber The reaction times for the humidity reading are shorter than in case of...

Страница 7: ...stments or similar are carried out the warranty services are free of charge but there is a charge for other services such as transport and packing costs Other claims especially those for damage occurr...

Страница 8: ...8052 Villingen Schwenningen Phone 49 0 7705 97 89 9 0 Fax 49 0 7705 97 89 9 20 info cs instruments com www cs instruments com Sales office NORTH Am Oxer 28c D 24955 Harrislee Phone 49 0 461 700 20 25...

Отзывы: