CRS AFS Purifier I
7
6. Thread the nuts, back ferrules and front ferrules onto
the tubing. Insert the tubing into the manifold fittings,
making sure that it reaches the bottom of the fitting.
7. Tighten the compression fittings. Compression fittings
normally require about ¾ turn past the finger-tight
position for a good seal. Do not overtighten or the
fittings may be damaged.
8. Remove the protective plastic cups from the manifold
check valves. Be careful, the ends of the check
valves are sharp!