Crowcon Tetra MO7237 Скачать руководство пользователя страница 22

Tetra gas test accessory kit

The Gas Test Accessory is a gas testing kit designed to enable gas testing of the 
Tetra multigas detector using a specially formulated, high stability long life quad 
gas mix. It can be used with Tetra units having sensors for Flammable, Oxygen, 
Carbon Monoxide and Hydrogen Sulphide gases, and all Tetra units with these 
sensors can be gas tested using the kit.

6.1 Gas testing 

Gas testing checks the sensor is responding within set limits to an applied gas 
of known composition. This can be performed as often as desired, but would 
usually be carried out each time the Tetra is issued for use. The Tetra itself will 
determine Pass/Fail status for the gas test.

In order to perform successful gas tests ensure:

•  The quad gas mixture used has the correct gas concentration, and that 

it is within the validity date specified by the supplier.

•  The gas flow path is leak tight. It is important to check that the flow 

plate is properly fitted to the Tetra, and the outlet tubing is not restrict-

ed in any way, nor additional tubing length used.

The Gas Test Accessory Kit is supplied in a convenient carry case and comprises; 
a gas cylinder containing the quad gas mix, a 'Trigger' regulator with intercon-
nect tubing, a magnet - used to activate Test mode, an aspirator plate to attach 
to the Tetra and a vent line. The Trigger regulator can be operated in two ways: 
(1) squeeze and hold - allows gas flow as long as the lever is held in, or (2) by 
lifting the lever - the flow is locked on. There are two versions of the kit; one 
for pumped Tetra units and one for non-pumped Tetra units supplied with an 
aspirator plate. Pumped Tetra units require the use of the plate with the integral 
bellows assembly.

6.2 How to perform a gas test

1.

  Ensure your Tetra unit is switched on and in normal operation.

2.

  Clip the flow plate onto the front of the unit and attach the hose from 

the Trigger regulator.  

Attach the outlet hose to ‘vent gas away’ - do not extend this hose 

and do not restrict or allow kinks. 

3.

  Swipe the magnet past the display adjacent to the LED lens. Your Tetra 

unit will activate the Gas Test and show ‘GAS TEST’ on the display.

4.

  Tetra will display a progress bar at the bottom, the names of the gas 

Gas testing 

Tetra

Содержание Tetra MO7237

Страница 1: ...Tetra Personal Multigas Monitor User Manual M07237 May 2010 Issue 10...

Страница 2: ...on performance standards EN50054 EN50057 EN61779 1 EN61779 4 EN50104 and EN50270 has been certified by Lloyd s Register 5 Repair of this equipment and gas sensor replace ment shall be carried out by t...

Страница 3: ...8 III Batteries 12 IV Alarm indications 14 V Fixing accessories 15 VI Flow sampling 16 VII Maintenance and calibration 21 VIII PC interface and software 23 IX i module replacement 24 X Specification...

Страница 4: ......

Страница 5: ...and aspirator bulb Optional spare battery pack for non rechargeable units Battery check The Tetra Personal Multigas Monitor has two battery options Li ion recharge able or alkaline non rechargeable b...

Страница 6: ...le steps to get your unit ready for use 1 Ensure the unit is in clean air 2 Switch on Press and hold the operator button until the red LED flashes The operator display screen will light up and the uni...

Страница 7: ...single click to confirm auto zero If the operator button is not pressed within the 10 second time out Tetra will proceed directly to Run mode without performing zero Tetra Quickstart guide LEDs Alarm...

Страница 8: ...any gas being monitored Tetra will activate the alarm signals Alarm signals The red and blue alarm LEDs will flash the sounder will emit a loud fast series of beeps the internal vibrator alarm will a...

Страница 9: ...tored in a safe non hazardous area 0 30 C 10 90 RH 4 Additional information For battery recharging information go to section III For fixing accessories go to section V For sampling section go to secti...

Страница 10: ...a single operator button and an intelligent user friendly display with automatic backlight Gas levels are continuously monitored provid ing normal gas readings peak readings and time weighted average...

Страница 11: ...ust tight to IP65 and with a non slip grip The internal structure has been carefully designed to make servicing easy and at the same time very rugged If the unit is dropped there will be no disruption...

Страница 12: ...B If auto zero fails a warning message will be displayed and an X will appear against the sensor that has failed Switch off To switch off the unit press and hold the operator button for 5 seconds A sh...

Страница 13: ...the icon is revolving 2 3 Display symbol guide Battery A full battery is represented by a battery icon showing a full six bars A low battery charge will show 1 to 2 bars When zero bars are shown the...

Страница 14: ...n the shaft rather than just a sampling tube Deselecting Peak mode clears stored peak information TWA display Shows the 15 minute or 8 hour time weighted average TWA for toxic gases monitored since la...

Страница 15: ...cessed using the IR communications link with Portables PC software See section VIII It can record data each minute for 50 hours up to a limit of 1500 events Tetra records the time and date for a numbe...

Страница 16: ...nsert the lead into the socket Switch on the power The unit would normally be left switched off for charging and will display a battery icon on the display sweeping from empty to full When charging is...

Страница 17: ...itable for the non rechargeable ver sion Energizer type LR6 MN1500 Gold Peak type 15A LR6 Duracell type MN1500 LR6 Varta type 4006 Alkaline batteries will typically last 11 hours To replace the batter...

Страница 18: ...rs are set to falling for deficiency monitoring Alarm latching Alarms can be set to be latched or unlatched Latched alarms will require the operator button to be pushed in order to clear the alarm Thi...

Страница 19: ...belt clip at the back of your Tetra unit The plate will lock automatically into place Create a chest harness by attaching one strap to the top connectors to go around the neck and the other to link ar...

Страница 20: ...onto the gas outlet nozzle 4 To remove the flow adaptor plate unscrew the thumbscrew and lift the plate away from the instrument 16 Gas in Gas out Flow adaptor The sampling tube supplied is normally...

Страница 21: ...ly stop Tetra will emit a rapid series of beeps and display a warning message To restart the pump clear the blockage and press the operator button Diffusion instruments When using the manual aspirator...

Страница 22: ...in any way nor additional tubing length used The Gas Test Accessory Kit is supplied in a convenient carry case and comprises a gas cylinder containing the quad gas mix a Trigger regulator with interc...

Страница 23: ...to confirm Calibration If the button confirmation for calibration is not made within 10 seconds then the process will revert to gas test as in 6 2 3 Apply calibration gas following step 4 in 6 2 Tetr...

Страница 24: ...se blocked or kinked Ensure no restriction to flow Tetra fails gas test Gas cylinder empty Check gauge replace cylinder as needed Gas cylinder out of date Check date and replace as needed Hose blocked...

Страница 25: ...ds as a minimum a monthly gas test to confirm sensor operation A test gas of known composition needs to be applied to verify sen sor response and alarm function Instrument calibration of all sensors s...

Страница 26: ...flow rate It is important that the flow rate is set correctly as otherwise pressure effects may distort calibration values and prevent the gas detector from working cor rectly To calibrate Crowcon po...

Страница 27: ...with access to reconfigure alarm levels opera tion run calibrations print reports and to access the event log Set up 1 Install Portables PC software on PC and install infrared adaptor Warning The infr...

Страница 28: ...screws as shown by point 6 Ease the top away from the body point 7 Press down lightly on the rubber seal protecting the sensor housing and slide forward to clear the internal chassis from the sensor...

Страница 29: ...mly and the sensor is still tightly located on the module board Replace the retaining ring on the clips 3 Attach the ribbon connector by squeezing the two retaining lugs on the module board toward eac...

Страница 30: ...r oxygen Repeatability 2 FSD 6 months Explosion protection Intrinsically Safe ATEX Essential Health and Safety Requirement clause 15 9 Safety certificate no BASEEFA03ATEX0193 Approval codes Europe ATE...

Страница 31: ...ure for Gas Test Accessory Kit 34 litre bottle 50 LEL methane 100ppm carbon monoxide 15ppm hydrogen sulphide 18 oxygen balance nitrogen Calibration gas contact Crowcon required gases depend on sensor...

Страница 32: ...irator hose please specify length in feet C03141 6 m Drop line C01245 Water trap Carrying and wearing C01842 Universal harness plate C01843 Shoulder strap C01844 Chest harness strap kit C01845 Carry c...

Страница 33: ...ea immediately Return instrument for recalibration or sensor replacement Autozero failed Zeroing in Switch off and restart in contaminated clean air atmosphere Cannot autozero due to Zeroing in Switch...

Страница 34: ...ted when the unit is exposed to an over range gas concentration there is a risk of damage to the pellistor sensor Restart should be carried out in a known fresh air environment Depleted oxygen levels...

Страница 35: ......

Страница 36: ...l 31 10 421 1232 Fax 31 10 421 0542 Email eu crowcon com Web site www crowcon com USA Office Crowcon Detection Instruments Ltd 21 Kenton Lands Road Erlanger Kentucky 41018 1845 USA Tel 1 859 957 1039...

Отзывы: