chemical toxic gas sensors may be degraded at
these levels.
Water should not be allowed to collect on the
sensor as this may impede gas diffusion. Use
with care in wet or humid environments, and
check response after use.
Persistent exposure to high levels of toxic gas
will shorten the life of the sensor. If the gas is
corrosive (e.g. hydrogen sulphide) damage may
occur to metal components.
Sensors may be cross sensitive to other gases.
If unsure, contact Crowcon or your local agent.
Use of high power radio transmitters in close
proximity to the instrument may exceed rated RFI
immunity levels and cause erroneous indications.
If such problems are experienced, remove
antennae to a reasonable separation from the
instrument (e.g. 30cm).
14. Accessories & Spare Parts
14.1 Accessories List
C01613
Gasman II Charger 12V DC input
C01628
Gasman II Charger with 230V UK
power supply
C01629
Gasman II Charger with 230V
European power supply
C01630
Gasman II Charger with 110V US
power supply
C01631
Gasman II Charger with 230V power
supply - no plug
C01632
Gasman II Charger with 110V power
supply - no plug
Gasman II
Instruction Manual
Issue 5
Page 28
11/99