Crowcon APM 135 Скачать руководство пользователя страница 2

MIN/MAX 

The MIN/MAX operation is used to record the 
maximum pressure level, minimum pressure level and 
the difference between the minimum and maximum 
pressure levels. 

The first press of the 

MIN/MAX

 button displays the 

maximum pressure recorded. The 

MAX

 icon is shown 

on the display 
The next press of the 

MIN/MAX

 button displays the 

minimum pressure recorded. The 

MIN

 icon is shown 

on the display. 
The next press of the 

MIN/MAX

 button displays the 

difference between the minimum and maximum 
pressure. The 

MAX-MIN

 icon is shown on the display. 

The next press of the 

MIN/MAX

 button returns to the 

standard pressure display. 

Pressing the 

ZERO

 button when either 

MAX

MIN

 or 

MAX-MIN

 displays are shown, only resets the current 

MIN and MAX values. It does not zero the current 
pressure reading. 

Example

Input pressure is 20 mbar, 

MIN

 mode selected then 

press 

ZERO

 button. 

MIN

 mode shows 20 mbar, 

MAX

 

mode shows 20 mbar, 

MAX-MIN

 mode shows 0 mbar. 

During 3 minute wait pressure initially increases to 
20.2 mbar then drops to 18.8 mbar. 

MIN

 mode shows 

18.8 mbar, 

MAX

 mode shows 20.2 mbar, 

MAX-MIN

 

mode shows 1.4 mbar. 

BATTERY REPLACEMENT

 

1. Remove the protective rubber boot by slipping it 

away from the bottom of the case and sliding it over 
the pressure spigots exposing the case back. 

2. Remove the battery cover from the bottom of the 

case by removing the four screws. 

3. Remove the three AA batteries and replace with 

good quality AA Alkaline batteries taking care to 
install them the right way round. 

4. Replace the battery cover carefully making sure it is 

alligned properly and replace the screws. Take care 
not to over tighten the screws otherwise it may 
damage the case. 

5. For best results and longer life always use quality 

alkaline batteries MN1500 or equivalent. 

 

CAUTIONS 

Exposure to silicone based products, corrosive gases, 
water / condensation and freezing should be avoided 
at all times as this will cause damage to the sensor 
and internal component corrosion. 

This instrument should not be stored or operated 
outside of its operating temperature range. 

Particular care must be taken not to exceed the 
maximum pressure ratings for the device as this 
may rupture the sensor membrane. This is not 
covered by the 

manufacturer’s warranty. 

NOTE: The differential/gauge instruments will only 
measure positive pressures i.e. positive pressure 
applied to positive port, or negative pressure applied 
to the negative port. 

WARRANTY 

This instrument is guaranteed against defects of 
workmanship and materials for a period of twelve 
months from date of invoice. 

During the warranty period a defective instrument will 
be repaired or replaced at the discretion of the 
manufacturer. This warranty does not cover damage 
or failure due to misuse or accident. Modification, 
adjustment or any alteration shall void the warranty. 

For any warranty claims to be considered the 
instrument must be returned to Anton along with proof 
of purchase at the senders cost. 

 

FCC Compliance 

NOTE: This equipment has been tested and found to 
comply with the limits for a Class A digital device, 
pursuant to part 15 of the FCC Rules. These limits are 
designed to provide reasonable protection against 
harmful interference when the equipment is operated 
in a commercial environment. This equipment 
generates, uses, and can radiate radio frequency 
energy and, if not installed and used in accordance 
with the instruction manual, may cause harmful 
interference to radio communications. Operation of 
this equipment in a residential area is likely to cause 
harmful interference in which case the user will be 
required to correct the interference at his own 
expense. 

 

 

STANDARD ACCESSORIES 

The APM 135 is supplied complete with: 

 

Three AA Alkaline batteries 

 

Operating instructions 

 

Protective rubber boot 

 

SPECIFICATIONS 

Power Supply 

3 x AA Alkaline Batteries 

Battery Life 

> 100 hours* 

Duty Cycle 

Continuous.  

Auto switch off after 5 mins. 

Operating 
Temperature 

-10 to +50 °C 

Weight 

360 grams including 
batteries 

Dimensions 

185 x 75 x 37 mm 

Accuracy 

±2% of reading 

* Using recommended Industrial batteries 
 

Units 

APM 135 

mbar 

-39.99 to +19.99 

+20.0 to +130.0 

PSI 

-0.29 to +1.89 

mmHg 

-15.00 to +19.99 

+20.0 to +98.1 

mmH

2

-199.9 to +199.9 

+200 to +1336 

inH

2

-8.03 to +52.63 

hPa 

-39.99 to +19.99 

+20.0 to +130.0 

Pa 

-400 to 9999 

kPa 

-1.99 to 13.00 

 

 

 

Отзывы: