APPENDIX H (ERROR INDICATOR SCREENS)
If certain malfunctions occur, MD50W and MD1 OOW series will display error messages:
oP
-Appears in display areas reserved for temperature readings when thermistor sensor
connected to that port is open circuited or not connected. Replace sensor with CPS model
number TMX3C.
SP
-Appears in display areas reserved for temperature readings when thermistor sensor
connected to that port is short circuited. Replace sensor with CPS model number TMX3C.
Err (Upper Left Display)
- Appears in upper left display when T function is selected and
one or both thermistor sensors are defective as indicated by oP or SP indicators mentioned
above.
Err (Lower Left Display)
-Appears in display areas reserved for temperature readings when
user tries to read saturation temperatures and calibration of pressure channels is invalid.
This defect requires manifold be returned to factory.
®
#73-156 RevB