CPI Canada Inc.
Installation
2
6.
Connect the ground wire to the chassis ground connector, and connect the mains power wires to
the terminals on the main fuse holder shown in figure 2-8. For 1 phase generators connect the
mains to F1 and F2; for 3 phase units connect the mains to L1, L2, and L3.
•
Ferrules should be used on the ends of the AC mains wires. These must be supplied by the
installer.
•
For China only, the power cable must be CCC approved.
2.6.5
Power Line Mains (Cont)
7.
Re-install the fan cover and reconnect the fan by reversing the previous steps.
THE FAN MUST BE CONNECTED AT ALL TIMES.
FAILURE TO CONNECT THE FAN WILL RESULT IN GENERATOR MALFUNCTION &
ERROR MESSAGES.
8.
Do not switch on mains power until requested to do so in a later step.
Figure 2-7: Termination of AC mains cable
AC Mains Termination Table
GROUND
L1
L2
L3
All 80 kW
Configurations
7 (178)
3.25 (83)
3.25 (83)
4.25 (108)
50 kW
208 – 230 VAC,
3P Configurations
7 (178)
5 (127)
4.75 (121)
5 (127)
40 kW & 50 kW
400 – 480 VAC,
3P Configurations
7 (178)
5.75 (146)
5.5 (140)
5.75 (146)
208 – 230 VAC
1P Configurations
7 (178)
5 (127)
5 (127)
N/A
*Dimensions are in inches (mm)
Use and disclosure is subject to the restrictions on the title page of this CPI document.
CMP 200
X-Ray Generator Service Manual Ch # 901476-02 Rev. P
Page 2-13