![COVVI AK140E316 Скачать руководство пользователя страница 74](http://html1.mh-extra.com/html/covvi/ak140e316/ak140e316_manual_2673968074.webp)
00 | Page 74
www.covvi.com | Unit 4, Quayside Business Park, Leeds, LS10 1DJ | 020 3949 9500
13.0 COVVI Go App
For Fast Open/Slow Close the following parameters can be established:
Fast Open / Slow Close Time -
A signal sent within this time is interpreted as an open signal and a signal sent
after this time is interpreted as a close signal.
For Single Site Alternating the following parameters can be defined:
Direction Change Time -
This is the time needed to pass for the signal to have a direction change. Any signal
sent within this time will keep the direction of the hand the same.
Open/Open Time -
This is the time in which a second open signal must be applied after a first open signal or
after the hand is fully opened for it to be interpreted as a grip switch trigger. This time must be shorter than
the direction change time for it to work.
Long Hold Open Time -
This is the time an open signal must be maintained after the hand is fully open for it
to be interpreted as a grip switch trigger.
Содержание AK140E316