![Covidien Newport e360 Скачать руководство пользователя страница 49](http://html1.mh-extra.com/html/covidien/newport-e360/newport-e360_service-manual_2673907049.webp)
Software Upgrade Procedure
Service Manual
2-25
14.
Confirm the software upgraded by checking the software version number (see
corner).
Figure 2-20.
Event History Log Screen
15.
Remove the USB flash drive from the USB port on the back panel of the ventilator.
2.10.3
Circuit Check Test and Diagnostic
WARNING:
Before returning the ventilator back for patient use after the software upgrade, you must perform this
procedure.
To do the circuit check and diagnostics, perform the following steps:
1.
Turn the ventilator off.
2.
Connect a medical-grade air source to the air inlet on the back panel of the ventilator (see
for air inlet location).
Содержание Newport e360
Страница 1: ...Service Manual Newport e360 Ventilator with Accessories TM...
Страница 7: ...v C Diagrams Index...
Страница 8: ...Page Left Intentionally Blank vi...
Страница 12: ...Page Left Intentionally Blank viii...
Страница 24: ...Introduction Page Left Intentionally Blank 1 12 Service Manual...
Страница 56: ...Maintenance Overhaul and Software Upgrade Page Left Intentionally Blank 2 32 Service Manual...
Страница 112: ...Calibration Procedures Page Left Intentionally Blank 5 16 Service Manual...
Страница 146: ...Ordering and Contact Information Page Left Intentionally Blank B 6 Service Manual...
Страница 148: ...Diagrams C 2 Service Manual Figure C 1 Newport e360 Ventilator Electrical Block Diagram...
Страница 149: ...Service Manual C 3 Figure C 2 Newport e360 Ventilator Pneumatic Diagram SPD2100A...
Страница 150: ...Diagrams Page Left Intentionally Blank C 4 Service Manual...
Страница 152: ...Page Left Intentionally Blank 1 2...
Страница 153: ......