Maintenance Procedures
Service Manual
2-7
Figure 2-3.
Exhalation Valve Adapter Replacement
To replace the exhalation valve adapter, refer to
and perform the following steps:
1.
Access the exhalation valve adapter by lifting the retaining latch and removing the exhalation valve
and exhalation flow sensor (see
and
2.
To remove the silicone exhalation valve adapter, use your finger to grasp the adapter and pull it
straight out.
3.
To replace the exhalation valve adapter, orient the adapter lip toward the inside of the exhalation valve
housing and press into place.
4.
Confirm that the adapter is properly seated.
5.
Reinstall the exhalation valve and exhalation flow sensor.
1
Retaining latch
2
Valve adapter
Содержание Newport e360
Страница 1: ...Service Manual Newport e360 Ventilator with Accessories TM...
Страница 7: ...v C Diagrams Index...
Страница 8: ...Page Left Intentionally Blank vi...
Страница 12: ...Page Left Intentionally Blank viii...
Страница 24: ...Introduction Page Left Intentionally Blank 1 12 Service Manual...
Страница 56: ...Maintenance Overhaul and Software Upgrade Page Left Intentionally Blank 2 32 Service Manual...
Страница 112: ...Calibration Procedures Page Left Intentionally Blank 5 16 Service Manual...
Страница 146: ...Ordering and Contact Information Page Left Intentionally Blank B 6 Service Manual...
Страница 148: ...Diagrams C 2 Service Manual Figure C 1 Newport e360 Ventilator Electrical Block Diagram...
Страница 149: ...Service Manual C 3 Figure C 2 Newport e360 Ventilator Pneumatic Diagram SPD2100A...
Страница 150: ...Diagrams Page Left Intentionally Blank C 4 Service Manual...
Страница 152: ...Page Left Intentionally Blank 1 2...
Страница 153: ......