Measured versus Calculated Saturation
Operator’s Manual
10-3
10.5
Measured versus Calculated Saturation
When calculating saturation from a blood gas partial pressure of oxygen (PO2), the
calculated value may differ from the SpO
2
measurement of a monitoring system.
This usually occurs when saturation calculations exclude corrections for the effects
of variables such as pH, temperature, the partial pressure of carbon dioxide (PCO2),
and 2,3-DPG, that shift the relationship between PO2 and SpO
2
.
Figure 10-1.
Oxyhemoglobin Dissociation Curve
functional saturation
%carboxyhemoglobin
fractional saturation
%methemoglobin
100
+
–
---------------------------------
100
=
1
% Saturation Axis
3
Increased pH; Decreased temperature, PCO2, and 2,3-DPG
2
PO2 (mmHg) Axis
4
Decreased pH; Increased temperature, PCO2, and 2,3-DPG