Front Panel Components Replacement
Service Manual
12-37
Single Board Computer (SBC) PCB Replacement
To remove the Single Board Computer (SBC) PCB
1.
Disassemble the monitoring system chassis. Reference
2.
Move the rear enclosure off to the side, retaining the front panel on a static-free
working surface.
3.
Remove and set aside the four T-6 torx screws.
4.
Gently disconnect both golden UFL antennae connectors for the antennae from
the board.
5.
Gently remove the SBC board, exerting even pressure on both sides.
Figure 12-21.
Single Board Computer (SBC) Removal
To replace the Single Board Computer (SBC) PCB
1.
Align the Single Board Computer (SBC) PCB connectors to the staggered connec-
tors on the Main PCB.
2.
Gently press down to connect the SBC PCB to the Main PCB, ensuring the board
is completely seated.
3.
Reconnect both golden UFL antennae connectors.
4.
Tighten the four T-6 torx screws to 3 inch-pounds.
5.
Realign the rear enclosure to the front panel on a static-free working surface.
Содержание Nellcor GR101704
Страница 1: ...Service Manual Nellcor Bedside Respiratory Patient Monitoring System...
Страница 8: ...Page Left Intentionally Blank vi...
Страница 32: ...Page Left Intentionally Blank 2 14 Service Manual Product Specifications...
Страница 46: ...Page Left Intentionally Blank 3 14 Service Manual Theory of Operations...
Страница 54: ...Page Left Intentionally Blank 4 8 Service Manual Product Overview...
Страница 96: ...Operation 6 34 Service Manual a Press an alternate menu option b Press EXIT AND RESTART to exit SERVICE MODE...
Страница 114: ...Page Left Intentionally Blank 7 18 Service Manual Trend Data Access...
Страница 120: ...Page Left Intentionally Blank 8 6 Service Manual Performance Considerations...
Страница 168: ...Page Left Intentionally Blank 10 46 Service Manual Modification and Testing...
Страница 192: ...Repair 12 4 Service Manual Figure 12 2 Exploded View of Internal Components...
Страница 232: ...Page Left Intentionally Blank 12 44 Service Manual Repair...
Страница 235: ......