Operation
6-28
Operator's Manual
Figure 6-4.
Technical Alarm Indicated
The following conditions are medium-priority technical alarms:
•
Preamp not connected: one or both preamplifiers are disconnected
•
Sensor not connected: one or more sensors are disconnected, or a reusable cable is disconnected
•
Poor signal: signal quality is too low for a reliable measurement; check sensor placement
•
Excessive light: an ambient light source is affecting the quality of the measurement; check sensor
placement, cover sensor if needed
•
Interference detected: an electromagnetic interference source such as an electrosurgical unit is affect-
ing the quality of the measurement; remove the interference or move the sensor
•
Replace sensor: both adult and infant/neonatal sensors are connected, or sensor is damaged or defec-
tive; make sure that only sensors appropriate for the patient are in use
•
Sensor unusable: the sensor has reached the end of its useful life (seen only on older monitors with
older preamplifiers; new monitors do not present this condition)
The following conditions are low-priority technical alarms:
•
Battery low: battery voltage is critically low, switch immediately to AC mains power to avoid an inter-
ruption in monitoring
•
USB storage error: there is no compatible USB storage device connected
•
Com port unavailable: the serial port device or its serial cable is disconnected
•
Printer unavailable: the printer or its serial cable is disconnected
•
Auto baseline set: 5 minutes of monitoring have elapsed without setting a baseline
Содержание INVOS 5100C
Страница 1: ...Operator s Manual INVOS Regional Saturation Patient Monitoring System Model 5100C...
Страница 12: ...Page Left Intentionally Blank x Operator s Manual...
Страница 26: ...Introduction Page Left Intentionally Blank 1 14 Operator s Manual...
Страница 30: ...Product Overview Page Left Intentionally Blank 2 4 Operator s Manual...
Страница 46: ...Installation Page Left Intentionally Blank 3 16 Operator s Manual...
Страница 50: ...Quick Setup Page Left Intentionally Blank 4 4 Operator s Manual...
Страница 142: ...Database Operational Mode and Area Under the Curve Page Left Intentionally Blank 8 24 Operator s Manual...
Страница 164: ...Specifications Page Left Intentionally Blank A 12 Operator s Manual...
Страница 170: ...Clinical Studies Page Left Intentionally Blank B 6 Operator s Manual...
Страница 176: ...Glossary Page Left Intentionally Blank C 6 Operator s Manual...
Страница 177: ......