Operation
6-20
Operator's Manual
The time scale may be adjusted in the Time scale menu or the User configuration menu. To
change the hours viewed (x-scale) of rSO
2
values plotted on the trend data graph, adjust the time
scale.
To adjust the time scale in the User configuration menu, press the HOME
key to display the
Main screen, then perform the following navigation bar options:
In the User configuration menu, select
NEXT
to scroll to
TIME SCALE
. Select the
INCREASE
and
DECREASE
options to adjust the time scale. The time scale range options are: 1, 2, 4, 8, 12, and 24
hours. The default time scale is 1 hour.
When finished, press the HOME
key to exit the User configuration menu and return to the
Main screen. On the Main screen, the time scale limit is numerically displayed.
6.5.7
USB Storage Rate
Note:
The USB flash drive should be inserted when the case begins to capture real-time data to the USB. Keep
the USB flash drive inserted for the full duration of the case. If removed, there will be gaps in the case data
collected. Regardless of USB flash drive usage, the INVOS™ system will monitor and store data to its internal
memory, so entire cases can be downloaded in the case history format (see section
).
USB flash drives not supplied by Covidien may not meet system requirements. Additional drives may be
ordered from Covidien (see section
Note:
For quicker access to parameters on page 2 of the User configuration menu, select
PREVIOUS
to wrap
around from the top to bottom of page 1, then select
NEXT
to access page 2.
BASELINE MENU
EVENT MARK
ALARM AUDIO ON/OFF
NEXT MENU
OUTPUT SELECT
USER CONFIGURATION
TIME SCALE
NEXT MENU
NEXT
PREVIOUS
OFF
ON
NEXT
PREVIOUS
INCREASE
DECREASE
NEXT
PREVIOUS
INCREASE
DECREASE
Содержание INVOS 5100C
Страница 1: ...Operator s Manual INVOS Regional Saturation Patient Monitoring System Model 5100C...
Страница 12: ...Page Left Intentionally Blank x Operator s Manual...
Страница 26: ...Introduction Page Left Intentionally Blank 1 14 Operator s Manual...
Страница 30: ...Product Overview Page Left Intentionally Blank 2 4 Operator s Manual...
Страница 46: ...Installation Page Left Intentionally Blank 3 16 Operator s Manual...
Страница 50: ...Quick Setup Page Left Intentionally Blank 4 4 Operator s Manual...
Страница 142: ...Database Operational Mode and Area Under the Curve Page Left Intentionally Blank 8 24 Operator s Manual...
Страница 164: ...Specifications Page Left Intentionally Blank A 12 Operator s Manual...
Страница 170: ...Clinical Studies Page Left Intentionally Blank B 6 Operator s Manual...
Страница 176: ...Glossary Page Left Intentionally Blank C 6 Operator s Manual...
Страница 177: ......