INVOS™ System Troubleshooting Chart
Indication
Possible cause
Action
Does not turn on
No AC power or battery power
Plug the INVOS™ system into a hospital-
grade electrical outlet.
Turn on the back panel AC power switch.
Blown fuse on line filter
Check the fuses.
Defective CPU board
Service the INVOS™ system.
1
Defective key panel
Service the INVOS™ system.
1
Blown fuse on internal module
Service the INVOS™ system.
1
Locks up/intermittent
Erratic AC power input
Change the electrical outlet.
Defective CPU board
Service the INVOS™ system.
1
No AC power or battery power
Plug the INVOS™ system into a hospital-
grade electrical outlet.
Turn on the back panel AC power switch.
Monitor turns off
Defective battery
Service the INVOS™ system.
1
Blown fuse on internal module
Service the INVOS™ system.
1
Defective fan
Service the INVOS™ system.
1
Key panel does not function
Defective key panel
Service the INVOS™ system.
1
INVOS™ system locked up
Reboot the INVOS™ system.
Defective internal module
Service the INVOS™ system.
1
No rSO
2
reading
See error messages in data channel area
on the display screen.
Follow the on-screen instructions.
Preamplifier not connected to the moni-
tor, sensor cable, or sensor
Connect the preamplifier, cable, or sensor.
Sensor not connected to the patient
Apply the sensor to the patient.
INVOS™ system monitor locked up
Reboot the INVOS™ system.
Defective CPU board
Service the INVOS™ system.
1
Defective preamplifier cable or connector
Service the INVOS™ system.
1
Blown fuse on the internal module
Service the INVOS™ system.
1
Defective preamplifier
Service the INVOS™ system.
1
Содержание INVOS 5100C
Страница 1: ...Operator s Manual INVOS Regional Saturation Patient Monitoring System Model 5100C...
Страница 12: ...Page Left Intentionally Blank x Operator s Manual...
Страница 26: ...Introduction Page Left Intentionally Blank 1 14 Operator s Manual...
Страница 30: ...Product Overview Page Left Intentionally Blank 2 4 Operator s Manual...
Страница 46: ...Installation Page Left Intentionally Blank 3 16 Operator s Manual...
Страница 50: ...Quick Setup Page Left Intentionally Blank 4 4 Operator s Manual...
Страница 142: ...Database Operational Mode and Area Under the Curve Page Left Intentionally Blank 8 24 Operator s Manual...
Страница 164: ...Specifications Page Left Intentionally Blank A 12 Operator s Manual...
Страница 170: ...Clinical Studies Page Left Intentionally Blank B 6 Operator s Manual...
Страница 176: ...Glossary Page Left Intentionally Blank C 6 Operator s Manual...
Страница 177: ......