IX. Specifications
Medical Electrical Equipment
Filac 3000 EZ/EZA Electronic Thermometer
Classified with respect to electrical shock, fire, and mechanical hazards only in accordance with UL60601-1
Classified with respect to electrical shock, fire, and mechanical hazards only in accordance with CAN/CSA C22.2 No. 60601-1-08
Dimensions (Approximate, without base):
in.- 3.2 W x 6.0 H x 2.1 D ; mm - 84 W x 152 H x 56 D
Dimensions (Approximate, with base):
in.- 4.1 W x 7.5 H x 3.1 D ; mm - 105 W x 190 H x 79 D
Weight (Approximate, without base):
13.5 Ounces; 385 Grams
Temperature Measurement Range:
30°C to 43°C; 86°F to 109°F
Ambient Operating Temperature Range:
10°C to 40°C; 50°F to 104°F; at 10% - 95% RH, non-condensing.
Average prediction times
(after insertion into measurement site):
Oral (EZ model):
3-5 seconds (non-fever temps),
8-10 seconds (fever temps)
Oral (EZA model):
6-10 seconds
Axillary Mode:
8-12 seconds
Rectal Mode:
10-14 seconds
Direct Mode (All Sites):
60-120 seconds
Typical times for switch to Direct Mode:
If no measurement site detected: 60 seconds
(after pulling probe from probe well)
If temperature does not stabilize: 70 seconds
(after measurement site insertion)
Pulse Timer:
60 Second count with a “beep” at 15 seconds, 2 “beeps” at 30 seconds, 1 “beep” at
45 seconds and 2 “beeps” at 60 seconds
Water Bath Accuracy
(35.5°C/95.9°F to 42.0°C/107.6°F):
Direct Mode:
+/-0.1°C - +/- 0.2°F
Standard Prediction Mode**:
+/-0.1°C - +/- 0.2°F
Quick Prediction Mode (Oral Only)**:
+/-0.3°C - +/- 0.5°F
**Greater than 95% of the prediction mode readings will be within the specified accuracy.
Patient Accuracy:
In standard predict mode, thermometer accuracy meets EN 12470-3
A Standard Prediction Mode reading and a Direct Mode reading will differ by
less than ±0.2°C (±0.4°F) on 98% of tested patients
Batteries:
Four “AA” Required. Standard IEC package size. Alkaline – 1.5 Volt
Approximately 6000 temperature readings.
Probe:
Replaceable.
For probe integrity/safety checks, refer to qualified Biomedical Technician or
return to Covidien.
Isolation Chamber:
Removable, replaceable, washable, Polycarbonate-polyester blend.