CorSolutions Version V2.003
SETTING UP THE PUMP
Hardware and Cable Connections.
The pump has been shipped with a Power Supply and appropriate
interface cables. The power supply and USB cable connect to the receptacles located on the rear panel of the
pump. The power supply connects to a 110-220V AC power source. The USB cable should be connected into
the host computer USB port. The hardware and cable connection set-up is now complete. The power switch
is also located on the rear panel of the pump.
If you are using two pumps with one computer, repeat the process.
IMPORTANT POINTS:
1) Pump is designed and calibrated for aqueous-based solutions. If other solution types are used,
the flow rates will not be accurate.
2) For proper operation, 1/32-inch tubing must be attached to the flow-sensor outlet. This tubing
is part of the dampening system and is needed for proper closed loop feedback for the flow
control. If a different outlet tubing type is desired, please employ a union at the end of the
1/32-inch outlet tube to connect to the desired tubing type. CorSolutions has adapters
available to convert from 1/32-inch to 1/16-inch (part number: Union-188-P)
3) The lines should be rinsed with 5% bleach before turning off the pump or after use with salty
or viscous solutions. This will ensure the flow sensor is maintained and will prevent it from
clogging and avoid the growth of bacteria.
4) Only use water-based solutions or 20/80 ration of Isopropynol/Water to rinse the lines. Use of
solvents such as acetone and Ethanol will cause damage to the pump tubing.