5. Charging the batteries on charging stand

  Put the pipet controller on the stand.
  Plug the charger into the stand outlet.
  Connect the charger to the adapter 

and to the main power.

6. Additional information

  Charging temperature: 10°C to 35°C.
  Before the pipet controller is activated for the first time, the batteries 

should be charged.

  Full charging time: 7 to 8 hours. 

  When the batteries are charged, the charging circuit disconnects 

automatically. 

  Service life of the batteries: Approximately 1,000 charging cycles, 

if used correctly. 

  Do not leave the pipet controller discharged for a long period of time.
  In case of a planned longer break in using the device, it is 

recommended to remove the batteries from the battery 
compartment or charge them every few months.