Photo 2
ATTENTION:
Please notice that you must enter the program as described under point 1.5.1 to change the
adjustments of the points 1.5.2. After you have stored the settings you change to the normal
program mode again by switch off the centrifuge for a short while.
1 PRODUCT DESCRIPTION
1.5.2 Call up of operating data
In the mode “Basic Adjustments” you can call up the operating data of the centrifuge.
Please proceed as described under point 1.5.1 to enter this program mode.
Press the key “accel/decel“(5). In the display “accel/decal” flashes the word “service”“.
With the adjusting knob (7) the different information can be called up:
A = previous starts of the centrifuge (see photo 3a)
H = previous operating hours
S = software version
E = list of previous error messages
F
U only for service purpose
P
C
The list of the last 99 error messages can be looked over by pressing the key „rpm/rcf“ (6) and leaf
through it by the adjusting knob (7). The respective error codes appear in the display „rpm/rcf“
(see photo 3b). Please look up in chapter 4.2.3 of this instruction manual for the relevant meanings.
The first two numbers (X) indicate the appeared errors ongoing from 00 to 99, the last two numbers (Y)
indicate the error code (i. e. 02 = imbalance sensor is defective).
Here as well you must shortly switch off the centrifuge for changing to the normal program mode again.
7
4
1
7
6
5
Y
X
CLS-EQ-104 October 2013 , Document 01, Revision 02