background image

Corning

®

 Lambda

 EliteMax  

Semi-automated Benchtop Pipettor

Quick Start Guide

Instrument Identification

Touch Screen 
Controller

Instrument Deck

Disposable Tips and 
Plates (not included)

Liquid Head 

Single-
channel 
Head

Multi-
channel 
Head

Manually 
Operated 
Tip Ejector

Unboxing

Tools needed:

 Phillips and flat head screwdrivers, 2 mm Allen 

wrench (provided), hammer, and needle-nose pliers.

 

Pry open top of box with the flat head screwdriver.

 

Remove the screws holding down the deck plate with the 
Phillips head screwdriver.

 

Remove the unit from the box, lifting from the bottom of the 
deck plate, and place on the benchtop.

 

Be sure to remove the holding bracket prior to use. The holding 
bracket is located next to the yellow warning label and can be 
removed using the Allen wrench provided.

Installation

1

2

4

1. 

Connect the Touch Screen Controller cable into the rear panel 

of the Corning Lambda EliteMax Benchtop Pipettor.

2. Connect the round connector of the universal external power 

supply to the rear panel of the instrument. Push in firmly until 
you feel it click into position.

3. Connect the AC power cord to the universal external power 

supply and then to the AC wall power receptacle. The AC 
power receptacle should be able to supply approx. 300W and 
MUST be grounded to a true earth ground.

4. Power up the unit by rocking the Reset/Off switch located in 

the back of the Corning Lambda EliteMax Benchtop Pipettor 

to the Reset position. The instrument should boot up, and the 
home screen will appear in a few seconds. 

For the complete Corning Lambda EliteMax Benchtop Pipettor Instruction 

Manual (CLS-AN-671DOC), visit 

www.corning.com/lifesciences

.