11
Instruction Manual
–
Cosmetics: Water containing tenside/Isopropyl alcohol
–
Food samples: Water containing tenside
–
Fuels: Water containing tenside
Before using other cleaning solutions than those listed above please check compatibility
with Corning Life Sciences prior to use. For safety, wear the proper protective gloves during
cleaning of the instrument.
NOTE:
If you require maintenance service, the instrument must be cleaned in advance to
avoid pollution of hazardous substances, and to be sent back in the original packing.
If the instrument will not be used for a long time, please turn Off and place in a
dry, clean, room temperature, and stable location.
Do not use cleanser within electrical ports.
Model
I-4010, I-4011, I-4012
Functions
Heating, cooling, and mixing
Temperature range
-15°C ambient to 100°C
Temperature setting range
0.1°C to 100°C
Control accuracy (°C at 20°C to 45°C)
±0.5°C
Uniformity (°C at 20°C to 45°C)
Max. ±0.5°C
Maximum heating rate
5.5°C/min.
Maximum cooling rate
5°C/min. (100°C to room temperature)
0.5°C/min. (below room temperature)
Speed range
200 to 1,500 rpm
Mixing diameter
3 mm
Display
LCD
Program
Up to 6 steps per 9 programs
Voltage
100 to 240V
Frequency
50/60Hz
Power
200W
Ambient operation range
+5°C to 40°C, up to 80% RH, non-condensing
The Axygen® Microtube shaker is designed to be safe when operated under the following
conditions:
Indoor use
Altitude up to 2,000 meters
Pollution Degree 2