background image

Axygen

®

 Microtube 

Shaker

Instruction Manual

Catalog Numbers: 
I-4010 
I-4011
I-4012

A Corning Brand

Содержание AXYGEN I-4010

Страница 1: ...Axygen Microtube Shaker Instruction Manual Catalog Numbers I 4010 I 4011 I 4012 A Corning Brand ...

Страница 2: ...on 5 5 1 Parameter Setting 5 5 2 Start and Stop 6 5 3 Running 6 5 3 1 Single Step Work Without Saving 6 5 3 2 Multiple Step Work Without Saving 7 5 3 3 Program with Saving mode 8 5 3 4 Load Program 8 5 3 5 Single Function Running 8 5 4 Temperature Calibration 9 5 5 Restore the Factory Settings 10 6 0 Trial Run 10 7 0 Troubleshooting 10 8 0 Maintenance and Cleaning 10 9 0 Limited Warranty 12 10 0 E...

Страница 3: ... power off before fitting accessories Temperature must always be set at least 50 C lower than the fire point of the media used Be aware of hazards due to Flammable materials or media with a low boiling temperature Overfilling of media Incorrect use of accessories Process pathogenic materials in closed vessels only Ensure the instrument and accessories comply with the safety and set up instructions...

Страница 4: ... in Section 1 100 mm Figure 1 Device placement 3 0 Unpacking Unpack the equipment carefully and check for any damages which may have arisen during transport before use Please keep original packaging Please contact the manufacturer supplier for technical support CAUTION If there is any apparent damage to the system please do not plug it into the power line Contact your local Corning support team 3 ...

Страница 5: ...L tubes 24 1 500 10 9 x 30 8 mm I 4010 D Heating block holds 5 mL tubes 8 1 500 13 6 x 42 mm I 4010 E Heating block cone bottom holds 5 mL tubes 8 1 400 17 x 51 5 mm I 4010 F Heating block holds 15 mL tubes 8 800 16 9 x 102 24 mm I 4010 G Heating block holds 50 mL tubes 4 600 29 6 x 105 95 mm I 4010 H Heating block for 0 2 mL tubes or 96 well PCR microplates 1 500 I 4010 I Heating block for 96 or ...

Страница 6: ...tachment ensure vessel type matches block type LCD screen Display to set experiment parameters Function button Switches the input parameters or programming Adjustable knob At parameter setting state rotate to alter function values Press to confirm setting During operations press to pause resume press and hold to stop end RS232 interface Used for temperature calibration Power interface Connects the...

Страница 7: ...n in rpm and will flash when selected Adjustable knob At parameter setting state rotate to alter function values Press to confirm setting During operations press to pause resume press and hold to stop end 5 0 Instrument Operation 5 1 Parameter Setting Press the function button The corresponding characters of the setting area will flash Rotate the parameter knob Input the target parameters for the ...

Страница 8: ...consistent with the temperature at the point of pausing Resume Press the knob once to resume to the Run while in the Pause phase Stop Press and hold the knob to stop the current Run 5 3 Running 5 3 1 Single Step Work Without Saving When powered On the LCD screen shows the parameters of the last run or factory settings Press the Temp Speed Time function buttons respectively the corresponding charac...

Страница 9: ...p is the first step by default Press the Prog function button to generate the second step The last line in the Set column now shows 1 2 NOTE Before setting the second step there is one step by default But after the Prog setting is complete the actual steps will be shown correctly once confirmed ...

Страница 10: ... program by a P1 P9 naming system Input the preferred P of the newly generated program and keep for your records then press the knob to complete the input The program saving is complete now Prog shows P1 5 2 means program number 1 has been selected P1 was saved with 5 steps and currently working through Step 2 5 3 4 Load Program While powered On press and hold the Prog button This enters the opera...

Страница 11: ...ture Calibration temperature Parameter adjustment knob Figure 3 Calibration interface 3 Rotate the Parameter knob to set the target temperature for calibration Press the parameter knob and confirm the set values 4 The actual Temp reaches the set temperature and remains constant for 20 to 30 minutes The calibration area will display the corresponding temperature on the display Record the measuremen...

Страница 12: ...arget parameters or programming Start working Observe the LCD display Stop working and power Off If these operations above are normal the device is ready to operate If not the device may be damaged during transportation please contact Corning Life Sciences for technical support 7 0 Troubleshooting Problem Solution Instruments will not power On Check whether the power line is unplugged Check whethe...

Страница 13: ... a dry clean room temperature and stable location Do not use cleanser within electrical ports Model I 4010 I 4011 I 4012 Functions Heating cooling and mixing Temperature range 15 C ambient to 100 C Temperature setting range 0 1 C to 100 C Control accuracy C at 20 C to 45 C 0 5 C Uniformity C at 20 C to 45 C Max 0 5 C Maximum heating rate 5 5 C min Maximum cooling rate 5 C min 100 C to room tempera...

Страница 14: ...or contact your local Corning sales office Corning Customer Service will help arrange local service where available or coordinate a return authorization number and shipping instructions Products received without proper authorization will be returned All items returned for service should be sent postage prepaid in the original packaging or other suitable carton padded to avoid damage Corning will n...

Страница 15: ...upport their claims We cannot make any claims or statements that our products are approved for diagnostic testing either directly or indirectly The customer is responsible for any testing validation and or regulatory submissions that may be required to support the safety and efficacy of their intended application For a listing of trademarks visit www corning com clstrademarks All other trademarks ...

Отзывы: