C O R N I N G   C E L L   C O U N T E R

Instruction Manual

WARNING! This count is below our optimum counting range (5.0 x 10

5

 cells/mL).

Figure 27. 

Warning message: not enough cells to count

WARNING! This count is above our optimum counting range (1.0 x 10

7

 cells/mL).

Figure 28. 

Warning message: too many cells to count

Bubbles

While looking at the images in the 
CytoSMART™ Cloud App, you may 
see bubbles (Figure 30). They are 
recognizable as big black circles. 
Bubbles can lead to false results, as they 
influence the distribution of cells and 
the volume inside the counting chamber. 
The cause for bubbles is often due to 
incorrect pipetting. Thus, make sure no 
air is included during pipetting a new 
sample.

Debris

When looking at the images from the 
CytoSMART Cloud App, you may see (cell) 
debris (Figure 31). This may be caused by 
not having properly cleaned the counting 
chamber. To resolve this, properly clean 
the counting chamber before loading a 
new sample.

Figure 29.

 Cell clump 

example

Figure 30. 

Air bubble 

example

Figure 31.

 Debris present 

in the counting chamber

No cells are  
visible 

This may be due to cell concentration of the sample being below  
5 x 10

5

 cells/mL, which is below the operating range of the Corning® 

Cell Counter (Figure 27). To resolve this, pellet the solution by 
centrifuging and resuspend the cells in a smaller volume. 
Another cause may be that the bulk sample solution has been left 
to stand for some time. Cells have then gravitated towards the 
bottom of the vial. By pipetting the suspension up and down before 
loading the counting chamber, you should obtain a homogenous cell 
suspension.

Too many cells 

This may be due to cell concentrations being above 1 x 10

7

 cells/mL  

(Figure 28). Dilute your cell suspension to obtain a lower concentration 
and count again.

Cell clumps

Cells can aggregate into clumps, leading 
to false results when analyzed (Figure 29). 
When cell clumps are visible, prepare a 
new sample. By gently resuspending the 
stock cell suspension, aggregates can be 
broken down.

Troubleshooting 

(Continued)

Содержание 6749

Страница 1: ...Corning Cell Counter Accurate Cell Counts in a Flash Instruction Manual Cat No 6749...

Страница 2: ...Plate Placement 12 Sample Loading 13 Live Dead Sample Loading 14 Counting Chamber Placement 14 4 0 Performing a Cell Count 15 Focusing of the Cells 15 Performing a Count 16 5 0 Connecting to the Clou...

Страница 3: ...the device 1 0 Introduction Technology to Make Your Work Easier Thank you for purchasing the Corning Cell Counter The Corning Cell Counter is a benchtop automated image based cell counter which uses...

Страница 4: ...ew of the Corning Cell Counter 1 Light Arc 2 USB Cable 5 Focus Knob 3 Optical Window 4 Adapter Plate 2 0 Getting Started Product contents include the following items Figure 2 1 Manual and Quick Starte...

Страница 5: ...on has completed an icon for the CytoSMART Cloud App will be visible on the desktop Figure 3 4 Start the CytoSMART Cloud App by double clicking the icon on the desktop A notification to connect the de...

Страница 6: ...can perform a single count or multiple counts within a project The CytoSMART Cloud App software automatically stores the results within a project which you can view later in the CytoSMART Connect Figu...

Страница 7: ...ownwards until it clicks into place on top of the Corning Cell Counter Figure 15 Figure 13 Start cell count Figure 14 Slide the adapter plate into place Figure 15 Carefully press the adapter plate dow...

Страница 8: ...installed you can place the counting chamber in the adapter plate that has been inserted on the Corning Cell Counter 4 0 Performing a Cell Count After placing the loaded counting chamber on the Corni...

Страница 9: ...nnect environment 5 If you want to perform another cell count simply press Start new cell count E Figure 19 Overview of the result page A B C D E 5 0 Connecting to the Cloud One of the features of the...

Страница 10: ...verview of all projects Figure 23 Navigating the portal is simple If you have multiple projects in an account use the Search Projects bar A to quickly search for a project To create a new project sele...

Страница 11: ...circles Experiment Pane The experiment pane provides an overview of the experiment i e project start date owner number of cells and viability Notes Pane Notes that you have added after counting the ce...

Страница 12: ...r before loading a new sample Figure 29 Cell clump example Figure 30 Air bubble example Figure 31 Debris present in the counting chamber No cells are visible This may be due to cell concentration of t...

Страница 13: ...causes not arising from defects in original material or workmanship This warranty does not cover motor brushes fuses light bulbs batteries or damage to paint or finish Claims for transit damage shoul...

Страница 14: ...te electrical and electronic equipment WEEE the Corning Cell Counter is marked with the crossed out wheeled bin and must not be disposed of with domestic waste Consequently the buyer shall follow the...

Страница 15: ...tact your local Corning sales office 2018 Corning Incorporated All rights reserved 2 18 CLS EQ 486 11319 30504 Corning Incorporated Life Sciences 836 North St Building 300 Suite 3401 Tewksbury MA 0187...

Отзывы: