8. 

Beeper (bp)

 

After Interval value is entered, bp is displayed and then On. Press UP or DOWN switch to select On. OFF, or 
FF, then press PRG to enter.  
 

ON- beeper sounds for Good, NG and capacity overload 

 

OFF- beeper sounds for overload only 

 

FF- beeper sounds for NG and capacity overload 

9. Units 

(Un) 

After the beeper is programmed, UN is displayed, then the current unit selection.  Press UP to cycle the units; 
lbf-in, ozf-in, kgf-cm, N-m and N-cm. 
 
After the units are entered, -S- is displayed to confirm programming completion and zero is shown. 
 
After High and Low setpoints are set and beeper is set to ON; 

 

In PP mode,  

 

Under LO setpoint  - no LED or beep 

 

Over LO setpoint – Green LED lights 

 

Over HI – red LED flashes and beeps sounds 

 

 

In PD mode, no indicator or beep for values before Peak Down is sensed.  

 

Under LO setpoint  - Red LED flashes and slow beep sounds 

 

Between LO and HI – Green LED lights and constant tone sounds 

 

Over HI – Red LED flashes and fast beep sounds 

 

 

In Real Time and C mode 

 

Under LO setpoint  - no LED light or beep 

 

Between LO and HI – Green LED lights and beep sounds 

 

Over HI – Red LED lights and beep sounds 

 
 

DATA MEMORY 

Storing and Recalling Data from Memory 

Memory functions work in PP and PD modes. Store up to 800 values in memory. 

 

1.  Measure torque in PP or PD mode, when the display resets by either Auto Zero or manually pressing CLR, 

the peak value is stored.  

2.  To recall a value, press MEM and the last stored memory is displayed. Press UP or DOWN to select a 

memory location and torque value. The display cycles between location and value.  

 

Clearing Data from Memory 

Single clear:

  Press MEM, then press UP or DOWN to move to a memory location. When the desired memory 

location is displayed, press CLR. Press CLR again to clear the data and ---- is displayed to confirm deletion. 

All clear:

  Press CLR and hold until the display shows ALL then press CLR again and ---- is displayed to confirm 

deletion. 

 

Downloading Memory Data 

1.  Press ON/OFF to turn on. 
2.  Press MEM and after memory data is displayed press MEM again and FA is displayed. Press UP or DOWN to 

select the first memory location then press MEM and LA is displayed.  Use UP or DOWN to select the last 
memory location. Press MEM again and the data is sent. While data is output –P- is displayed.  

NOTE:  by pressing DOWN switch for more than 1 second, the download function can be terminated.  
 

Statistics 

Press Statistics and the number of data, Max, Min, and Avg are displayed in sequence.  