Section 5

Menu Function

12

5

CooperSurgical, Inc.

Tn-t/Tn-b

The Tn-t/Tn-b option is for 

calibration

 of the 

displayed temperature 

of the table top (Tn-t) and the 

drawer (Tn-b). If there is an offset between the value on the display and any measurements made 

with a high precision external temperature sensor, this can be corrected. The new temperature 

reading will be kept as the displayed value with temperature controlling conducted on this basis. 
Follow these steps to navigate in the Tn-t/Tn-b option.

Action

Key

Display

Press and hold UP+DOWN key for 3 seconds to 

enter the menu.
The menu “UNIT” will appear in the display.

Press DOWN key until the option “Tn-t/Tn-b” 

appears in the display.

Press and hold set-point key to change values.

While holding the set-point key in, press either the 

arrow UP or arrow DOWN key to select the value 

of temperature.
When the required temperature has been selected 

let go of the setpoint key.

Press and hold UP+DOWN key for 3 seconds to exit 

the menu.

Caution: Any change in the factory set calibration should always be based on very certain 

temperature measurements. CooperSurgical only recommend the use of a high quality PT-100 

sensor that is calibrated with the precision instrument used for measuring according to the 

manufacturers specifications. Ensure an optimal contact between the sensor and the place 

measured. Use only a sensor type that is correct for the purpose, and which has the correct 

size. If in doubt contact your local CooperSurgical distributor.

INT.T

The INT.T (Integral Time) option is for changing the 

base value

 for the 

PID controller

. This should 

not be attempted by unauthorised persons. If set at a different level the controlling principle will 

be affected.
From the factory it is set to a closely calculated value specific for the model.

tn-t

CAUTION: Do not change the value here. If by mistake any changes are made or 

doubts occur if the value are correct please set the REST function in the menu for 

the restore of the defaults.

Содержание K-Systems R65

Страница 1: ...79 02 02 sales coopersurgical com fertility coopersurgical com CooperSurgical Inc 95 Corporate Drive Trumbull CT 06611 USA Document K33118UM 2 DRF 5570 24 November 2020 1639 K Systems R65 Trolley User...

Страница 2: ......

Страница 3: ...eneral Description 4 SECTION 4 BASIC OPERATION 5 Features and Operation 5 Assembling the R65 5 Connection to the Mains 5 Start of Apparatus 5 Keyboard Functions 5 Operating the Heated Area 5 Temperatu...

Страница 4: ...onic equipment 20 SECTION 7 TROUBLESHOOTING 21 SECTION 8 TECHNICAL DATA 22 SECTION 9 WARRANTY INFORMATION AND LIMITS ON LIABILITY 23 SECTION 10 RETURNING PRODUCT 24 Returned Goods Policy 24 Customer S...

Страница 5: ...erSurgical Inc products should not hesitate to contact us if there are any unclear points or ambiguities in this manual CooperSurgical Inc Phone 1 203 601 9818 International Phone 45 46 79 02 02 sales...

Страница 6: ...e instructions for use for important cautionary information such as warnings and precautions that cannot for a variety of reasons be presented on the medical device itself Manufacturer Date of manufac...

Страница 7: ...er code 1212807 Optional Equipment Optionalequipmentconsistsofarangeofwarmingblocks PleasecontactyourlocalCooperSurgical distributor or CooperSurgical Inc for details about the full range of warming b...

Страница 8: ...g assisted reproductive technology ART treatments General Description The stainless steel surface maintains a preset temperature in the range ambient to max 42 C The temperature is electronically cont...

Страница 9: ...upply using the enclosed PSU Start of Apparatus To power up the unit press the Heat ON OFF button on the keyboard The display indicates the current temperature either on the table plate or in the draw...

Страница 10: ...se or lower the set temperature When the desired temperature is reached let go of the set point key Note The set point can be adjusted to a reading between ambient and 49 9 C Combination Keys Action K...

Страница 11: ...en the temperature stabilizes to the set point level Warming Up Warming up from 20 0 C to 37 0 C takes less than 20 minutes under normal conditions When a cold item is placed on the heated area it may...

Страница 12: ...The first option UNIT will appear in the display Press UP DOWN key to navigate within the Menu While holding the set point key in press either the arrow UP or arrow DOWN key to change values within th...

Страница 13: ...ooperSurgical Inc Overview of Options S 1 S 2 Send Version 3 C F UNIT On O RS232 0 05 0 05 0 05 Tn t 0 1 0 1 INT T M 1 H 1 Ti St M 1 H 1 St St On O HEAT On O A St On O HOUR REST Software version Calib...

Страница 14: ...ction Key Display Press and hold UP DOWN key for 3 seconds to enter the menu The menu UNIT will appear in the display Press and hold set point key to change values While holding the set point key in p...

Страница 15: ...o select between ON and OFF OFF The RS232 function is turned off ON Data is send once every two seconds When the required setting is activated let go of the set point key Press and hold UP DOWN key fo...

Страница 16: ...ed let go of the setpoint key Press and hold UP DOWN key for 3 seconds to exit the menu Caution Any change in the factory set calibration should always be based on very certain temperature measurement...

Страница 17: ...s and hold UP DOWN key for 3 seconds to enter the menu The option UNIT will appear in the display Press DOWN key until the option Ti St appears in the display Press and hold set point key to change va...

Страница 18: ...d hold set point key to change values While holding the set point key in press either the arrow UP to change minutes or arrow DOWN key to change the hour When the time is set let go of the set point k...

Страница 19: ...splay Press and hold UP DOWN key for 3 seconds to enter the menu The first menu UNIT will appear in the display Press DOWN key until the option HEAT appears in the display Press and hold set point key...

Страница 20: ...UP DOWN key for 3 seconds to enter the menu The first menu UNIT will appear in the display Press DOWN key until the option A st appears in the display Press and hold set point key to change values Wh...

Страница 21: ...FF When the required setting is activated let go of the set point key Press and hold UP DOWN key for 3 seconds to exit the menu If you want to see the time on the display when the heat is turned on Fo...

Страница 22: ...gate in the REST option Action Key Display Press and hold UP DOWN key for 3 seconds to enter the menu The option UNIT will appear in the display Press DOWN key until the option REST appears in the dis...

Страница 23: ...visibly soiled 3 Allow to air dry 4 Visually inspect for the absence or presence of remaining soil While inspecting give particular attention to verifying soil has been removed from the hard to clean...

Страница 24: ...and electronic equipment WEEE Environmental implications WEEE contains materials that are potentially hazardous to the environment and to human health Therefore when this instrument has reached its e...

Страница 25: ...n 0 5 C from set temperature Check the set temperature Wait for the temperature to stabilize No heating Set point below ambient Change set point e g 37 C Display remains off when HEAT key is activated...

Страница 26: ...sions WxDxH 664x581x760mm Table plate 661x493mm Warming area 450x380mm Temperature range Ambient to 49 9 C Temperature accuracy 0 3 C Temperature set read Digital LED display Power consumption 240W He...

Страница 27: ...misuse improper care improper use of chemicals or cleaning methods loss theft servicing by non authorized personnel or negligent or intentional conduct on the part of the owner or user of the workstat...

Страница 28: ...products must be cleaned and sterilized prior to shipment A signed decontamination declaration may be required All products should be carefully and adequately packed preferably in original packaging R...

Страница 29: ...gical Inc 10 Customer Service Contact Details Tel 45 46 79 02 02 Fax 45 46 79 03 02 E mail sales coopersurgical com fertility coopersurgical com US Only Customers Contact Details Tel 800 243 2974 Fax...

Страница 30: ...79 02 02 Fax 45 46 79 03 02 New Purchases E mail sales coopersurgical com Service Related E mail service origio com fertility coopersurgical com US Service Representatives CooperSurgical Inc 95 Corpor...

Страница 31: ......

Страница 32: ......

Отзывы: