Clearing Data from Memory
1.
Single clear:
When a memory value is displayed, press CLR/PRG, then “CLR” displays and blinks 3 times.
Press CLR/PRG switch again before “CLR” stops blinking and the memory value is erased.
2.
All clear:
Press CLR/PRG and DOWN simultaneously until “ALL” is displayed. While “ALL” is blinking,
press CLR/PRG again. Display changes to “CLR” and blinks 3 times. Press CLR/PRG switch again before
“CLR” stops blinking and all memory data is erased.
Downloading Memory Data
1. Press ON/OFF to turn on.
2. Press UP and DOWN switches simultaneously. The display blinks “FA” twice and then displays numbers
with decimal points in between (.0.0.1 for first memory location).
3. Press UP or DOWN switch to select the desired beginnin memory location and press CLR/PRG to enter.
The display blinks “LA” for the ending memory location.
4. Press UP switch or DOWN switch to select the desired ending memory location and press CLR/PRG to
enter.
5. Press CLR/PRG switch to download the data. While downloading the data, the display shows “-P-“.
Note:
By pressing DOWN switch for more than 1 sec, the download function can be terminated.
Auto Power Off
To maximize the life of the battery, power automatically shuts off after 10 minutes of non-use.
IMPORTANT!
Use the provided AC adapter/charger exclusively and plug into the correct AC output. It takes 8
hours to fully recharge. Do not recharge for more than 12 hours. When fully charged, disconnect the AC
adapter/charger is to avoid overcharging.
In order to avoid high heat, explosion or toxic fumes, please note the following precautions:
1. Be sure to use only the provided AC charger exclusively.
2. Do not plug the charger into unspecified higher voltage.
3. Do not recharge until LOBAT icon appears.
System Reset
When battery power is completely depleted, the tester may not work even though it has been recharged. In this
case, press the System Reset Button.
RS-232 Serial Output
Data output is available in the PD, PP and C modes. In PD and PP modes peak data is output when ZERO is
pressed or activated by the AUTO ZERO function. In C mode, the gauge outputs data continuously.
RS-232C Signal: 8 data, 2 stop, no parity. Baud rate: 9600 bps.
Peak Data Output Format
[CAN] _ _ _ [SO] [value] _ [SI] [unit] [CR]
[CAN]:
ASCII control code 24
_:
Space (code 32)
[SO]:
ASCII control code 14
[value]:
Output data with sign and decimal point. Plus sign represents CW torque and minus
sign for CCW. [value] always occupies six locations and empty locations will be filled
with spaces.
[SI]:
ASCII control code 15
[unit]:
N*m _ _ _=N·m
kgf*cm=kg·cm
lb*in=lb·in
[CR]:
ASCII control code 13 (Carriage Return)
Continuous Output Data Format
[CAN][value][CR]