Table 4 Guidance and MANUFACTURER’s declaration – electromagnetic IMMUNITY –

for ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING 

Guidance and manufacturer’s declaration – electromagnetic immunity

IMMUNITY test

3 Vrms

3 V/m 

Compliance 

level

IEC 60601 

TEST LEVEL

Conducted RF

IEC 61000-4-6 

Radiated RF 

IEC 61000-4-3

3 V/m

80 MHz to

2.5 GHz 

3 Vrms

150 kHz to

80 MHz 

Electromagnetic environment - 

guidance

Portable and mobile RF communications 

equipment should be used no closer to 

any part of the device, including 

cables, than the recommended 

separation distance calculated from the 

equation applicable to the frequency of 

the transmitter.

Recommended separation distance

The device is intended for use in the electromagnetic environment specified below.

The customer or the user of the device should assure that it is used in such an environment.

where

 P

 is the maximum output power

rating of the transmitter in watts (W) 

according to the transmitter manufacturer

and 

d

 is the recommended separation

distance in metres (m).

Field strengths from fixed RF 

transmitters, as determined by an 

electromagnetic site survey,   should be

less than the compliance level in each

frequency range.

a

b

Interference may occur in the vicinity of 

equipment marked with the following 

symbol:

P

=

d

P

=

d

 80 MHz to 800 MHz 

P

=

d

2.333

  800 MHz to 2,5 GHz 

1.167 

NOTE 1 

At 80 MHz and 800 MHz, the higher frequency range applies. 

NOTE 2 

These guidelines may not apply in all situations. Electromagnetic propagation is affected by 

                    absorption and reflection from structures, objects and people.

a

Field strengths from fixed transmitters, such as base stations for radio (cellular / cordless) telephones 

and land mobile radios, amateur radio, AM and FM radio broad-cast and TV broadcast cannot be 

predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF 

transmitters, an electromagnetic site survey should be considered. If the measured field strength in 

the location in which the device is used exceeds the applicable RF compliance level above, the 

device should be observed to verify normal operation. If abnormal performance is observed, 

additional measures may be necessary, such as re-orienting or relocating the device.

Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3V/m.

b

1.167 

 

150 kHz to 80 MHz 

80 MHz to 800 MHz 

800 MHz to 2.5 GHz 

Table 6 Recommended separation distances between portable and mobile 

RF communications equipment and the ME EQUIPMENT or ME SYSTEM –

for ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING 

Recommended separation distances 

between  portable and mobile RF communications equipment and the device.

The device is intended for use in an electromagnetic environment in which 

radiated RF disturbances are controlled. The customer or the user of the device

can help prevent electromagnetic interference by maintaining a minimum distance 

between portable and mobile RF communications equipment (transmittters) and the 

device as recommended below, according to the maximum output power of the 

communications equipment.

Rated maximum output

power of transmitter

(W) 

Separation distance according to frequency of transmitter (m)

For transmitters rated at a maximum output power not listed above, the recommended 

separation distance 

d

 in metres (m) can be estimated using the equation applicable to 

the frequency of the transmitter, where 

P

 is the maximum output power rating of the 

transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1    At 80MHz and 800MHz, the separation distance for the higher frequency range applies.
NOTE 2    These guidelines may not apply in all situations. Electromagnetic propagation is affected by 

                 absorption and reflection from structures, objects and people.

0.01

0.1

1

10

100

0.117

0.117
0.369
1.167
3.690
11.67

0.233
0.738
2.333
7.378
23.33

P

=

d

P

=

d

1.167

2.333

P

=

d

0.369
1.167
3.690
11.67

1.167

30

31

Содержание WHT1-808-B

Страница 1: ...n case you have problems Thank you very much for selecting COOEY Blood Pressure Monitor WHT1 808 B GUANGDONG COOEY MEDICAL ELECTRONICS CO LTD Arm Type Version 1 0 EC REP Zone A 5 F Investment Building...

Страница 2: ...the day Why the blood pressure I get from the hospital is different from home If the result is the same if measuring on the right arm TROUBLESHOOTING 24 SPECIFICATIONS 25 ATHORIZED COMPONENT 26 CONTAC...

Страница 3: ...D Symbol for COMPLIES WITH MDD 93 42 EEC REQUIREMENTS Symbol for MANUFACTURER Symbol for SERIAL NUMBER Symbol for TYPE BF APPLIED PARTS Symbol for DIRECT CURRENT Symbol for ENVIRONMENT PROTECTION Wast...

Страница 4: ...fetus are unknown Manufacturer will make available on request circuit diagrams component parts list etc This unit is not suitable for continuous monitoring during medical emergencies or operations Ot...

Страница 5: ...sonel If measured three times per day and the battery is fully charged it can be used for about 20 days 3 Storge and use the blood pressure monitor at the cool dry and ventilated environment Avoid to...

Страница 6: ...OU START 2 As pictured in the right the blinking numeral representing HOUR Press Query button to change the numeral Each press will increase the numeral by one in a cycling manner 3 Press User 1 butto...

Страница 7: ...T BEFORE YOU START 5 Repeat step 2 and 3 to confirm MONTH DAY and YEAR 6 Repeat step 2 and 3 to confirm the measurement unit 7 After confirming the measurement unit the LCD will display dOnE and the m...

Страница 8: ...n the floor back and arm supported The central of the cuff should maintain at the same level as the right atrium of the heart Pair up the Blood Pressure Monitor with Your Device 1 Turn on Bluetooth an...

Страница 9: ...inish the whole measurement press another user button the blood monitor will begin measuring again B Maximum 60 records are both for user 1 and user 2 1 With WHT1 808 B successfully pair up with your...

Страница 10: ...ry records 4 When in the memory mode press the User 1 button to recall the measurement history of User 1 or press the User 2 button to recall the measurement history 5 When no history stored for the s...

Страница 11: ...tton for 3 seconds to clear the memory 2 When the LCD display dEL ALL press Query button to confirm 3 The LCD will display dEL dOnE and then shut off 4 If you wish to stop clearing the memory you may...

Страница 12: ...ccuracy if the measurement is taken in the following circumstances Maintenance 20 21 INFORMATION FOR USER INFORMATION FOR USER To obtain the best performance please follow below instructions Put in a...

Страница 13: ...vere CAUTION 1 Individual blood pressure varies every in one day it is also affected by the way you tie your cuff and the your measurement position so please take the measurement at the same condition...

Страница 14: ...White Backlight V A 86 1mm L x24mm W Measurement mode Oscillographic testing mode Measurement range Accuracy Normal working condition Storage transportation condition Measurement perimeter of the uppe...

Страница 15: ...ent Part 1 General requirements for basic safety and essential performance IEC EN 60601 1 11 2010 Medical electrical equipment Part 1 11 General requirements for basic safety and essential performance...

Страница 16: ...nd ME SYSTEMS IEC 60601 test level Compliance level Electrostatic discharge ESD IEC 61000 4 2 Electromagnetic environment guidance Floors should be wood concrete or ceramic tile If floors are covered...

Страница 17: ...RF transmitters an electromagnetic site survey should be considered If the measured field strength in the location in which the device is used exceeds the applicable RF compliance level above the devi...

Отзывы: