GEN1000-GE Operator Manual
10 November 8 2019 | 291180-ENG R00
3.1.3 Water Supply
Table 3-1
Water Supply Parameters
Parameter
Measurement
Connectivity
¼” Quick Connect
Flow
0.26 gallons / hour (1 Liter / hour) reverse osmosis
Pressure
Max.
– 115 psi (7.9 bar), Min. – 5 psi (0.3 bar)
pH
7.0 ± 0.5
Filtration
< 0.00008 inch (2 microns)
Resistance
0.01 to 0.02 Megaohm-
cm (MΩ-cm)
Conductivity
2.0 to 0.2 μS
3.1.4 Condensate Drain
A 3/4
″ (19 mm) drain is provided underneath the chamber, located near the back. The drain may
be extended to a nearby floor drain, as required. If there is no floor drain nearby, the optional
condensate pan may be used to collect the condensate and the optional condensate pump may
be used to pump the condensate to the drain.
3.2 Remove the Chamber from the Shipping Pallet
The GEN1000 chamber is designed to be removed from the shipping pallet with a fork lift. In
smaller facilities without access to a fork lift, the shipping pallet may be ordered with an optional
built-in ramp to unload the chamber.
To remove the chamber from the pallet with a fork lift:
1.
Remove the metal shipping bracket (Figure 3-1) from both sides of the pallet.
2.
Position the forks of the fork lift truck as shown in Figure 3-2.
Figure 3-1
Remove the L-Brackets
Figure 3-2
Position the Forks
Failure to use a water source with the quality stated in Table 3-1 will void the product
warranty.
Содержание GEN1000-GE
Страница 1: ...GEN1000 GE OPERATOR MANUAL 291180 ENG R00...
Страница 12: ...GEN1000 GE Operator Manual 4 November 8 2019 291180 ENG R00...
Страница 16: ...GEN1000 GE Operator Manual 8 November 8 2019 291180 ENG R00...
Страница 26: ...GEN1000 GE Operator Manual 18 November 8 2019 291180 ENG R00...
Страница 34: ...GEN1000 GE Operator Manual 26 November 8 2019 291180 ENG R00...
Страница 46: ...GEN1000 GE Operator Manual 38 November 8 2019 291180 ENG R00...
Страница 53: ......