Operation
26
October 2018 | 287849-ENG, R00
5.9.3 Closed Loop Dimmable Lighting System
Lamp canopies incorporate dimmable ballasts. Automatic adjustment of the light intensity can be
set within the programmed range for as low as 10% for fluorescent, 10% for LED, or 30% with HID
lamps, to the maximum intensity. Incandescent lamps are controlled in light levels.
Prior to operating dimmable fluorescent lamps, run the lamps at full intensity for a period of 100
hours to burn off impurities on the filament ends of the lamps left during the manufacturing
process.
5.10 Additive Humidification
5.10.1 Air-assisted Spray Nozzle Humidification (ASNH)
This optional additive humidification uses the pressure from the water supply to atomise the water
through a small orifice in the spray nozzle. The spray nozzles are located within the air handling
system of the chamber and the water is directly injected into the air stream before it is introduced
to the chamber.
A minimum of 50 psi (3.5 bar) is required to operate spray nozzles. If this water pressure is not
available, the CPSNH option is required to achieve the required pressure.
A 3/8
″ compression OD quick-connect fitting is supplied on the chamber for connection to the
building’s purified water supply.
The humidification system needs reverse osmosis filter
Failure to burn-in the fluorescent lamps before dimming will significantly reduce the
life of the tubes.
Do not add potentially irritating disinfectants or anticorrosive compounds to the water.
Do not use well water, industrial water, cooling circuit water, or any potentially
chemically or bacteriologically contaminated water.