Stopping the VFC400-USB
VFC400-USB User Guide
49
Stopping the VFC400-USB
You can configure a VFC400-USB so it can be stopped with the
START
/CLEAR/STOP
button.
This feature is useful when you take the logger out of a shipment and don’t want to
falsify the statistics with readings taken after the shipment completion. The stop
function is enabled in the
tab during configuration. When enabled,
following will stop the unit:
Press and hold the
START
/CLEAR/STOP
button.
STOPPED
is shown in addition to the
symbol.
Keep holding the button until the
symbol
disappears, then release the button within 2
seconds.
STOPPED
is shown permanently on the display and
the logger will no longer take readings.
The last recorded temperature is now no longer
shown, but minimum and maximum trip values
remain.
The logger will
not
stop if you
l
release the button before the
symbol
disappears; or
l
keep holding the button after
STOPPED
remains permanently on until the
symbol reappears.
Once the logger has stopped, the last alarm status before the logger stopped will
continue to be indicated with a cross or a tick symbol on the display, for each