background image

2.  To adjust the temperature alarm set point, press the 

or 

 button. 

(Do 

not press the trigger while adjusting the alarm set points). 

Press and hold 

to advance the display rapidly. 

 
To turn the Hi/Low Alarm function On/Off: 
1.  Locate the SET ALARM ON/OFF switch in the handle. 
2.  Slide the switch to the desired position. 
 
With the Alarm function set to ON, (indicated by ((( HI and LOW ))) appearing 
on the display), while taking temperature measurements, an alarm will sound 
when the temperature measured is above or below the High and Low alarm 
set points that have been entered. If the temperature is above the High 
Alarm set point, ((( HI will flash on the display. If the temperature is below the 
Low Alarm set point, LOW ))) will flash on the display. 
The alarm will sound regardless of the Display Mode if the temperature 
measured is above or below the High and Low alarm set points that have 
been entered. 
 

LOCK FUNCTIONS 

The lock function allows the thermometer to take long-term measurements 
without having to hold the trigger. When the lock function is activated, the 
unit will take continuous temperature measurements until the LOCK function 
is set to OFF. 
 
To use the lock function: 
1. Locate the LOCK ON/OFF switch in the handle. 
2. Slide the switch to the ON position. 
3. Close the handle. 
4. Press and release the trigger to begin taking measurements. 
 
The unit will continue to take measurements without holding the trigger until 
the LOCK ON/OFF switch is set to OFF. 
 
5. When finished taking measurements, locate the LOCK ON/OFF switch in 

the handle. 

Traceable

®

 is a registered trademark of Control Company   

©

2008  Control Company.  92-4484-00 Rev. 1 052108 

6. Slide the switch to the OFF position. 
7. Close the handle. 

 
EMISSIVITY 

 

Emissivity adjustment is optional. Emissivity adjustments are used to provide 
a truer temperature reading. The emissivity value appears at the top of the 
display (

ε

=#.##). 

Different materials radiate infrared energy at slightly different temperatures. 
The emissivity adjustment is used to compensate for different types of 
materials. The default emissivity of 0.95 will cover 90% of typical 
applications. 

The emissivity table provides a guide of different emissivity values for 
different materials.  
 
When the emissivity of an object is unknown, use a non-infrared 
thermometer, such as a thermometer with a surface probe to measure the 
object’s surface temperature. Adjust the emissivity until the temperature of 
the infrared thermometer matches the temperature of the surface probe. The 
emissivity value arrived at by this method may be used to measure similar 
materials. 
 

To adjust the emissivity (optional): 

1.  Press the MODE button until 

EMS

 appears on the display. 

2.  To adjust the emissivity value, press the   or   button. 

(Do not press the 

trigger while adjusting the emissivity value). 

Press and hold to advance 

the display rapidly. 

 
The emissivity can be set from 0.10 to 1.00. This emissivity value will be 
used for all temperature measurements until the value is reset. 

 
DISPLAY MESSAGES 

“1”

 will appear on the display when the temperature being measured is 

outside the range of the unit. 

 
ALL OPERATIONAL DIFFICULTIES 

If this thermometer does not function properly for any reason, replace the 
battery with a new high-quality battery (see Battery Replacement section). 
Low battery power can occasionally cause any number of “apparent” 
operational difficulties. Replacing the battery with a new fresh battery will 
solve most difficulties. 

 
BATTERY REPLACEMENT 

Erratic readings, faint readings, no display, or a battery symbol appearing on 
the display are all indications that the battery must be replaced. 

Locate the indentions on either side of the plastic 
case near the trigger. Open the battery compartment 
by pulling the rubber portion of the handle out and 
away from the trigger. 

Pull hard, it requires a good 

amount of force to open the battery compartment.

 

Remove the exhausted battery and replace it with a 
new 9-volt alkaline battery. Close the battery cover 
by snapping it shut.

  

 

 
 

 
 
 
WARRANTY, SERVICE, OR RECALIBRATION 

For warranty, service, or recalibration, contact: 

CONTROL COMPANY 

4455 Rex Road 

Friendswood, Texas 77546 USA 

Ph. 281-482-1714   Fax 281-482-9448 

E-mail [email protected] 

www.control3.com 

 

 

Control Company is ISO 9001 Quality-

Certified by DNV and ISO 17025 accredited 

as a Calibration Laboratory 

by A2LA.

 

Отзывы: