Installation & Maintenance
10
515 BR03 - 15 August 2021
Electrical Connection
Rear Panel Connections
Note: Depending on the specification and age of the instrument,
there are different versions of the upper option card. Ensure the instrument
is wired as per the correct terminal designation printed on the rear of the
instrument bezel and below.
Figure 3 shows the connections on the original rear panel of the instrument
where relays 1 to 4 (terminals 32-35) share a single Relay Common
(terminal 31). Note: The 5 way relay terminal block is GREEN.
Figure 3 Rear Panel Connections - Original
On the newer rear panel option cards the instrument relays 1 and 2
(terminals 32-33) share Relay Common 1-2 (terminal 31) and relays 3 and
4 (terminals 34-35) share Relay Common 3-4 (terminal 36).
Note: On these option cards the 6 way relay terminal block is ORANGE.
Figure 4 shows the new option card with the RS-485 port (terminals 19-21).
Figure 4 Rear Panel Connections - New RS-485 Version