D-10
Appendix D --- Error codes
System Error:
100-50198
Description:
Incorrect camera board.
Corrective Action.
Try the following:
■
Please check all camera board types..
■
Update the firmware
■
Run SCANtest, test 9 See Appendix A
■
Replace the CIS Element
■
If CIS element 1&2 or 3& or 5&6 is affected replace CIS Unit
■
Run SCANtest, test 9 See Appendix A
■
Replace the Scanner Controller SUI
System Error:
100-50199
Description:
Incorrect Main Electronics Board and combination Camera Board.
Corrective Action:
Try the following:
■
Please validate the combination for this scanner.
■
Check that all the cables are connected correctly to the Cameras Boards.
■
Update the firmware
■
Run SCANtest, test 9 See Appendix A
■
Replace the CIS Unit
■
Run SCANtest, test 9 See Appendix A
■
Replace the Scanner Controller SUI.
System Error:
100-50231
Description:
Firmware download is in progress
Corrective Action
.
Try the following:
System Error:
100-50232
Description:
Firmware is incomplete.
Corrective Action.
Try the following:
■
Please download new firmware
System Error:
100-50234
Description:
Scanner is in safe mode.
Corrective Action.
Try the following:
■
Reboot the system
System Error:
100-50235
Description:
Scan subsystem failed to initialize (FW/SW mismatch)
Corrective Action.
Try the following:
■
Reload system
Содержание XD2490
Страница 1: ...Technical Service Guide for XD2490 FSM TSG 002 1 1 ...
Страница 6: ...vi ...
Страница 10: ...2 2 Design MG52D Block Diagram ...
Страница 11: ... 2 3 Design MG52D Wiring Diagram ...
Страница 20: ...4 2 Part replacement MG52D 4 2 Part replacement MG52D Identifying parts ...
Страница 36: ...A 6 Appendix A ...