background image

 

 

 

 

 

PC-HELPER 

 

 

High-Resolution Analog Input Board 
for Low Profile PCI 

AD16-64(LPCI)LA 

User’s Guide 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

CONTEC CO., LTD.

 

Содержание AD16-64(LPCI)LA

Страница 1: ...PC HELPER High Resolution Analog Input Board for Low Profile PCI AD16 64 LPCI LA User s Guide CONTEC CO LTD ...

Страница 2: ...r missing items contact your retailer Product Configuration List Board AD16 64 LPCI LA 1 First step guide 1 Disk 1 API PAC W32 1 Standard sized bracket 1 Warranty Certificate 1 Serial Number Label 1 1 The bundled disk contains the driver software and User s Guide this guide Board First step guide First step guide Serial Number Label Warranty Certificate Warranty Certificate Disk API PAC W32 Standa...

Страница 3: ...ange without notice All relevant issues have been considered in the preparation of this document Should you notice an omission or any questionable item in this document please feel free to notify CONTEC CO LTD Regardless of the foregoing statement CONTEC assumes no responsibility for any errors that may appear in this document or for results obtained by the user as a result of using this product T...

Страница 4: ...ecautions 6 Environment 7 Inspection 7 Storage 7 Disposal 7 2 SETUP 9 What is Setup 9 Using the Board under Windows Using the Driver Library API PAC W32 9 Using the Board under Windows Using Software Other than the Driver Library API PAC W32 9 Using the Board under an OS Other than Windows 10 Step 1 Installing the Software 11 Which Driver to Use 11 Starting the Install Program 12 Step 2 Setting th...

Страница 5: ... O signals Counter signals and Control signals Connection 39 4 FUNCTIONS 41 Analog Input Function 41 1 Setting the Conversion Conditions 42 2 Starting Stopping Operation 49 3 Monitoring the Status and Acquiring Data 50 4 Reset 53 Counter Function 54 1 Setting the Operating Conditions 54 2 Starting Stopping Operation 55 3 Monitoring the Status and Acquiring Data 55 4 Reset 56 Digital Input Function...

Страница 6: ...s 68 Uninstalling the driver 68 6 ABOUT HARDWARE 69 For detailed technical information 69 Hardware specification 69 Block Diagram 71 Control Signal Timings 72 Control Signal Timings for Analog Input 72 Control Signal Timings for Analog Output 73 About Calibration 74 ...

Страница 7: ...vi AD16 64 LPCI LA ...

Страница 8: ...els The selection of the single ended input and the difference input can be set with software Rich set of basic functions Compact system providing high precision analog inputs Equipped with analog inputs analog input control signals TTL level 3ch digital inputs TTL level 4ch digital outputs TTL level 4ch and counter TTL level 32 bit 1ch Substantial control functions Capable of analog input either ...

Страница 9: ...ovided For more details on the supported OS applicable language and new information please visit the CONTEC s Web site Data Logger Software C LOGGER Stored on the bundled disk driver library API PAC W32 C LOGGER is a data logger software program compatible with our analog I O products This program enables the graph display of recorded signal data zoom observation file saving and dynamic transfer t...

Страница 10: ...erminal M3 x 68P EPD 68A 2 3 4 Screw Terminal M3 x 96P EPD 96A 1 3 4 Screw Terminal M3 5 x 96P EPD 96 1 3 BNC Terminal Unit analog input 32ch ATP 32F 1 3 BNC Terminal Unit analog input 8ch ATP 8 1 3 5 1 ADC 68M 96F optional cable is required separately 2 PCB68PS 0 5P or PCB68PS 1 5P optional cable is required separately 3 Two sets of cables are required to use both connector CNA and CNB 4 Spring u...

Страница 11: ...freight prepaid with a copy of the original invoice and a Return Merchandise Authorization to the distributor or the CONTEC group office from which it was purchased This warranty is not applicable for scratches or normal wear but only for the electronic circuitry and original boards The warranty is not applicable if the device has been tampered with or damaged through abuse mistreatment neglect or...

Страница 12: ...destruction of equipment and resources Understand the meanings of these labels to operate the equipment safely DANGER indicates an imminently hazardous situation which if not avoided will result in death or serious injury WARNING indicates a potentially hazardous situation which if not avoided could result in death or serious injury CAUTION indicates a potentially hazardous situation which if not ...

Страница 13: ... failure Be sure that the personal computer power is turned off Do not install or remove the board to or from the slot while the computer s power is turned on Otherwise the board may malfunction overheat or cause a failure Be sure that the personal computer or the I O expansion unit power is turned off Make sure that your PC or expansion unit can supply ample power to all the boards installed Insu...

Страница 14: ...ing this product keep it in its original packing form 1 Put the board in the storage bag 2 Wrap it in the packing material then put it in the box 3 Store the package at room temperature at a place free from direct sunlight moisture shock vibration magnetism and static electricity Disposal When disposing of the product follow the disposal procedures stipulated under the relevant laws and municipal ...

Страница 15: ...1 Before Using the Product 8 AD16 64 LPCI LA ...

Страница 16: ...are and hardware You can use the diagnosis program later to check whether the software and hardware function normally Step 1 Installing the Software Step 2 Setting the Hardware Step 3 Installing the Hardware Step 4 Initializing the Software Step 5 Checking Operations with the Diagnosis Program If Setup fails to be performed normally see the Setup Troubleshooting section at the end of this chapter ...

Страница 17: ...of this manual This chapter Step 2 Setting the Hardware Chapter 3 External Connection Chapter 5 About Software Chapter 6 About Hardware For using the board under an OS other than Windows and Linux see the following parts of this manual This chapter Step 2 Setting the Hardware Chapter 3 External Connection Chapter 6 About Hardware ...

Страница 18: ...ystem as Windows XP Although some user interfaces are different depending on the OS used the basic procedure is the same Which Driver to Use CONTEC has two analog I O drivers API AIO WDM and API AIO 98 PC API AIO WDM is a new driver for analog I O under Windows This driver was developed to be easier to use and to provide additional functions above those provided by the previous API AIO 98 PC drive...

Страница 19: ...PAC W32 on your PC 2 The API PAC W32 Installer window appears automatically If the panel does not appear run drive letter AUTORUN exe 3 Click on the Install the drivers button Before installing the software in Windows XP 2000 or NT log in as a user with administrator privileges CAUTION ...

Страница 20: ... WDM Analog I O driver 3 Click on the Install button Clicking the Details button displays detailed information about API AIO WDM and API AIO 98 PC Run the installation 1 Complete the installation by following the instructions on the screen 2 The Readme file appears when the installation is complete You have now finished installing the software ...

Страница 21: ...the factory defaults untouched You can change board settings later Replacing the Bracket This board is shipped with a Low Profile size bracket mounted To plug the board into a standard size slot replace the bracket with the bundled standard size bracket The replacing method is as follows Figure 2 1 Replacing the Bracket Screws Remove the screws and replace it with the Standard size bracket Standar...

Страница 22: ...f the boards to distinguish them The board IDs can be set from 0 Fh to identify up to sixteen boards If only one board is used the original factory setting Board ID 0 should be used Setting Procedure To set the board ID use the rotary switch on the board Turn the SW1 knob to set the board ID as shown below Figure 2 3 Board ID Settings SW1 SW1 BOARD ID SW1 BOARD ID SW1 0 1 2 3 4 5 6 7 9 A B C D E F...

Страница 23: ...our hands Otherwise the board may malfunction overheat or cause a failure If the terminals are touched by someone s hands clean the terminals with industrial alcohol Do not install or remove the board to or from the slot while the computer s power is turned on Otherwise the board may malfunction overheat or cause a failure Doing so could cause trouble Be sure that the personal computer or the I O ...

Страница 24: ...rd can be allocated Before attempting to install the board first determine what PC resources are free to use The resources used by each board do not depend on the location of the PCI bus slot or the board itself If you remove two or more boards that have already been installed and then remount one of them on the computer it is unknown that which one of the sets of resources previously assigned to ...

Страница 25: ...etup information INF file to register the board Source folder The setup information INF file is contained in the following folder on the bundled disk INF WDM AIO You have now finished installing the hardware INF WDM AIO The name of the board you have just added is displayed AD16 64 LPCI LA ...

Страница 26: ...From My Computer Control Panel select System and then select the Device Manager tab You can also open Device Manager by right clicking on My Computer and selecting Properties 2 The installed hardware appears under the CONTEC Devices node Open the CONTEC Devices node and select the device you want to setup the device name should appear highlighted Click Properties The name of the board you have jus...

Страница 27: ... name you set here is used later when programming The initial device name that appears is a default value You can use this default name if you wish Make sure that you do not use the same name for more than one device You have now finished installing the initial setting of Software The name of the board you have just added is displayed AD16 64 LPCI LA ...

Страница 28: ...t status Check Method To check the analog input data connect to an external signal source The figure below shows an example of checking by connecting to an external signal The example below is for analog input channel 0 on the AD16 64 LPCI LA Wiring Diagram Analog input Figure 2 4 Wiring Diagram Input data remains indeterminate when no input pin is connected The input pin for the channel not conne...

Страница 29: ...tarting the Diagnosis Program Click the Diagnosis button on the device property page to start the diagnosis program The name of the board you have just added is displayed AD16 64 LPCI LA The name of the board you have just added is displayed AD16 64 LPCI LA ...

Страница 30: ...O The upper row of circular lamps indicates the digital input states Red indicates the bit is ON and brown indicates OFF Clicking the lower row of switches turns the digital output bits ON or OFF Counter input Selecting a counter channel displays the count value and state of that counter channel Clicking the zero clear button resets the count to zero ...

Страница 31: ...nd displays the file for you to view Clicking Diagnosis Report prompts you to specify where to save the report text file 2 The diagnosis report contains the following data Version of OS Device Information File Information Initialization interrupts current input or output state for each channel The name of the board you have just added is displayed AD16 64 LPCI LA ...

Страница 32: ...battery or similar if you do not have any other suitable signal source Also check that a prescribed voltage can input the channel by the short circuit as an analog ground The diagnostic program works correctly but the application program does not The diagnostic program uses the API TOOL functions If the diagnostic program works correctly other applications should work correctly also If you have a ...

Страница 33: ...2 Setup 26 AD16 64 LPCI LA ...

Страница 34: ...B68PS P ADC 68M 96F or PCA68PS P is used to connect the board to external devices The cable is used together with a terminal block to connect external devices Two sets of cables are required depending on the number of channels used Please refer to chapter 1 for more information on the supported cable and accessories Figure 3 1 Interface Connector Shape CNA CNB Interface Connector CNA CNB Connector...

Страница 35: ...ed channel 0 7 Ο Ο Ο Only CNB is used channel 32 39 5 CNA B is used 3 channel 0 7 32 39 5 Ο 4 Ο 4 Ο 4 1 AI External Start Trigger Input AI External StopTrigger Input AI External Clock Trigger Input 2 Counter Gate Control Input Counter Up Clock Input Counter Output 3 Two sets of terminal blocks and optional cables are required each 4 Make wiring on the CAN side 5 Two or more only of channel 32 39 s...

Страница 36: ...put 43 Analog Ground for AI 18 52 Analog Ground for AI Analog Input 58 50 16 Analog Input 42 Analog Input 12 19 53 Analog Input 28 Analog Input 57 49 15 Analog Input 41 Analog Input 13 20 54 Analog Input 29 Analog Input 56 48 14 Analog Input 40 Analog Input 14 21 55 Analog Input 30 Analog Ground for AI 47 13 Analog Ground for AI Analog Input 15 22 56 Analog Input 31 Analog Input 55 46 12 Analog In...

Страница 37: ...e Control Input Gate control input signal for counter Counter Up Clock Input Count up clock input signal for counter Counter Output Count output signal Digital Ground Common digital ground for digital I O signals external trigger inputs Reserved Reserved pin N C No connection to this pin Figure 3 3 Pin Assignment of interface connector Single Ended Input Do not connect any of the outputs and power...

Страница 38: ...og Input 46 B26 A26 Analog Input 38 Analog Input 23 A23 B23 Analog Input 31 N C B27 A27 N C Analog Ground for AI A22 B22 Analog Ground for AI N C B28 A28 N C Analog Ground for AI A21 B21 Analog Ground for AI Analog Input 61 B29 A29 Analog Input 53 N C A20 B20 N C Analog Input 45 B30 A30 Analog Input 37 N C A19 B19 N C Analog Input 60 B31 A31 Analog Input 52 Digital Input 00 A18 B18 Digital Output ...

Страница 39: ...e Control Input Gate control input signal for counter Counter Up Clock Input Count up clock input signal for counter Counter Output Count output signal Digital Ground Common digital ground for digital I O signals external trigger inputs Reserved Reserved pin N C No connection to this pin Figure 3 4 Pin Assignment of interface connector Single Ended Input Do not connect any of the outputs and power...

Страница 40: ...ut 23 57 Input Control External Sampling Stop Trigger Input Analog Input 22 45 11 Analog Input 22 Input Control External Sampling Clock Input 24 58 Digital Ground Analog Input 21 44 10 Analog Input 21 N C 25 59 N C Analog Input 20 43 9 Analog Input 20 N C 26 60 N C Analog Ground for AI 42 8 Analog Ground for AI N C 27 61 Digital Ground Analog Input 19 41 7 Analog Input 19 N C 28 62 N C Analog Inpu...

Страница 41: ...ther connect outputs to each other Doing either can result in a fault If analog and digital ground are shorted together noise on the digital signals may affect the analog signals Accordingly analog and digital ground should be separated Leave Reserved pins unconnected Connecting these pins may cause a fault in the board CAUTION ...

Страница 42: ... Input 30 B26 A26 Analog Input 22 Analog Input 07 A23 B23 Analog Input 15 N C B27 A27 N C Analog Ground for AI A22 B22 Analog Ground for AI N C B28 A28 N C Analog Ground for AI A21 B21 Analog Ground for AI Analog Input 29 B29 A29 Analog Input 21 N C A20 B20 N C Analog Input 29 B30 A30 Analog Input 21 N C A19 B19 N C Analog Input 28 B31 A31 Analog Input 20 Digital Input 00 A18 B18 Digital Output 00...

Страница 43: ...e Control Input Gate control input signal for counter Counter Up Clock Input Count up clock input signal for counter Counter Output Count output signal Digital Ground Common digital ground for digital I O signals external trigger inputs Reserved Reserved pin N C No connection to this pin Figure 3 6 Pin Assignment of interface connector Differential Input Do not connect any of the outputs and power...

Страница 44: ... may effect the cross talk noise between channels If the board and the signal source receive noise or the distance between the board and the signal source is too long data may not be input properly An input analog signal should not exceed the maximum input voltage relate to the board analog ground If it exceeds the maximum voltage the board may be damaged Connect all the unused analog input channe...

Страница 45: ...is not connected the conversion data is not determined If the board and the signal source receive noise or the distance between the board and the signal source is too long data may not be input properly An input analog signal should not exceed the maximum input voltage relate to the board analog ground If it exceeds the maximum voltage the board may be damaged Connect all the unused analog input c...

Страница 46: ...bles the external clock input to the counter You can use this function to control the external clock input to the counter The external clock input to the counter is enabled when the input is High and disabled when the input is Low As the pin has an internal pull up on the board or card the default if not connected is High As a result the external clock for the counter is enabled if this pin is not...

Страница 47: ...3 External Connection 40 AD16 64 LPCI LA ...

Страница 48: ...solution and stores it in memory You can set a variety of conditions for analog input including the input channel sampling period and sampling start stop conditions Analog input processes are classified as follows 1 Setting the Conversion Condititions Event Status Sampling Repeat Data aquisition Start Stop 2 Starting Stopping Operation 3 Monitoring the Status and Acquiring Data Status Memory 4 Res...

Страница 49: ...ion is 16 bit Input Mode Input Mode indicates the method of connecting analog input signals The input modes available are single ended input and differential input The single ended input mode is suitable for the environment in which the potential difference between the signal source and ground and noise components can be ignored For the environment in which they cannot be ignored the differential ...

Страница 50: ... conversion is performed consecutively starting from channel 0 Software setup is not required as this board uses a fixed channel conversion priority Range Range means the range of voltages at which analog input can be performed Software setup of the range is not required as this board uses a fixed range of voltages AD16 64 LPCI LA 10V ...

Страница 51: ...on the device itself or in the driver The device buffer can operate as FIFO or ring memory The application calls an API function at an appropriate timing and fetches the conversion data from the device buffer The device buffer mode provides function that allows the number of items of conversion data using the number of sampling times as a unit to obtain the number of items of conversion data direc...

Страница 52: ...h the memory has stored a fixed amount of data or in which the memory has become full The FIFO memory is used to obtain all input data from analog input in a short or infinite period of time Ring format In the ring format the memory contains storage areas arranged in a ring Input data items are written to the memory sequentially When it stores data exceeding the limit it overwrites the area storin...

Страница 53: ... in which case the conversion is terminated by the software abort command Input data items are stored to the memory sequentially The repetition state can be subject to status monitoring and application notification Clock The sampling clock controls the sampling frequency You can select both the internal sampling clock and external sampling clock Internal sampling clock The clock signal from the on...

Страница 54: ...vel comparison condition is satisfied in the rising direction The start condition is satisfied when the analog signal at the specified channel passes the comparison level in the rising direction Input data items are stored to memory starting with those at solid dots The above sketch shows that the level comparison condition is satisfied in the falling direction The start condition is satisfied whe...

Страница 55: ... condition is satisfied in the rising direction The stop condition is satisfied when the analog signal at the specified channel passes the comparison level in the rising direction Input data items are stored to memory ending until those at solid dots The above sketch shows that the level comparison condition is satisfied in the falling direction The stop condition is satisfied when the analog sign...

Страница 56: ...conversion start condition is satisfied The event is nullified when the conversion start condition is software Repeat end event This even occurs whenever a repetition is completed End of device operation event This event occurs when the entire operation including repetitions is completed Stored specified sampling times event This event occurs when sampling has been performed for the number of time...

Страница 57: ...mpling data stored This status is set to ON when input data stored in memory has reached the amount corresponding to the preset number of times of sampling If the memory format is FIFO the status is set to OFF when the amount of input data in the memory falls below the value corresponding to the preset number of times of sampling as data is acquired Once the status is set to ON when the memory for...

Страница 58: ...s acquired differently depending on the memory format used Data acquisition in FIFO format When FIFO memory is used the oldest data is always read first The following sketch shows an image of data acquisition in FIFO format When data is acquired from the memory the free memory space increases by that data size When data is acquired next the oldest one of the existing data items is taken from the m...

Страница 59: ...lowing sketch shows an image of data acquisition in ring format The sampling count obtained is always the number of times of sampling for up to the latest data shaded portion below The larger the number of samples taken the older the data item acquired first As the ring memory retains data even after that data is acquired you can fetch the same data any number of times ...

Страница 60: ...V 0 Ex When input data 49152 is input at a resolution of 16 bit in the 10 V range Voltage 49152 x 10 10 65536 10 5 0 4 Reset Various states can be reset by executing the following reset commands Status This command resets the sampling clock error status and AD conversion error status Memory This can only be used when the transfer mode is set to device buffer mode This command resets the following ...

Страница 61: ...an example of using the compare count load function After the counter starts the first count match occurs when the count reaches 1000 When the count reaches 1000 the counter value at which the second compare count match is to occur 2000 is set This continues with the next value from the array being set each time a count match occurs After the final value from the array is loaded operation can star...

Страница 62: ...ounter is running Status The current state of the device can be checked by obtaining the device status The following types of device status are available Counter operating The device operating status is ON from the time the operation start command is executed until operation stops due to a stop command or error Compare count match The compare count match status turns ON when a count match occurs a...

Страница 63: ...A 4 Reset Various states can be reset by executing the following reset commands Counter reset Resets the counter This restores the counter to its state after power on Status Resets the compare count match status and overrun status ...

Страница 64: ...n Bytes Individual input bits can be input in byte units When the number of input points of the device is 4 the individual input bits are arranged as shown below and the byte data to be input is a value between 0 and 15 depending on the states of the bits EX Input of bit 3 OFF bit 2 ON bit 1 OFF bit 0 ON Byte data 05 5H Digital filter A digital filter can be used on the input bits The filter time ...

Страница 65: ...put bit can be changed to ON or OFF by specifying the bit and setting it to 1 or 0 Output in Bytes Individual output bits can be output in byte units When the number of output points of the device is 4 the individual output bits are arranged as shown below and byte data to be output is a value between 0 and 15 Ex Output of bit 3 ON bit 2 OFF bit 1 ON bit 0 OFF Byte data 10 AH Bit 3 Bit 2 Bit 1 Bit...

Страница 66: ...u html Version information on each driver English APIPAC Each installer AIO DISK1 DISK2 DISKN AioWdm CNT DIO HELP HELP file Aio Cnt INF Each INF file for OS WDM Win2000 Win95 linux Linux driver file cnt dio Readme Readme file for each driver Release Driver file on each API TOOL API_NT For creation of a user specific install program API_W95 UsersGuide Hardware User s Guide PDF files ...

Страница 67: ...ut sampling buffer memory usage and interrupt events such as occurrences of errors Driver option check using a demo driver even without the board installed For details refer to the help file The help file provides various items of information such as Function Reference Sample Programs Tutorial FAQs and Troubleshooting Use them for program development and troubleshooting Accessing the Help File 1 C...

Страница 68: ...eference when writing your own programs To use the sample programs specify the device name in the property page for the program The sample programs are stored in Program Files CONTEC API PAC W32 AIOWDM Samples Running a Sample Program 1 Click on the Start button on the Windows taskbar 2 From the Start Menu select Programs CONTEC API PAC W32 AIOWDM SAMPLE 3 A sample program is invoked ...

Страница 69: ...og input using a callback routine Analog output Simple sample program SingleAo Perform single analog output from specified channel MultiAo Perform single analog output from multiple channels Device buffer Ao Perform standard analog output using a FIFO buffer AoPoll Perform standard analog output by polling AoEx Perform analog output for multiple channels using a FIFO buffer AoLong Perform long dur...

Страница 70: ... diagnostic program Procedure 1 Chose the measure device from device list 2 Click the button written with the function name to measure the execution speed of the function Please choose from a list the number of channels used for conversion in function AioMultiAi and AioMultiAo Input the transmission data size in function AioGetAiSamplingData and AioSetAoSamplingData The transmission data is set by...

Страница 71: ...a of the memory is acquired Data in the FIFO memory can be confirmed visually The number of channels used the internal external clock the conversion speed and the sampling frequency at which an event generates can be set Since the notification of a sampling clock error event is sent please make use of it for the conversion spec measurement under various conversion conditions 1k The name of the boa...

Страница 72: ... is conversion data taken in the memory This can be visually checked in a memory image Event generation sampling frequency When the number of input sampling in FIFO reaches this frequency the event generates Total input sampling frequency It is the total number of samplings for application in the memory Measurement may stop by the following errors Sampling clock error It means that the conversion ...

Страница 73: ... taskbar From the Start Menu select Settings Control Panel 2 Double click on Add or Remove Programs in the Control Panel 3 Select CONTEC API AIO WDM driver and CONTEC API AIO WDM VerX XX Development environment from the list of applications Click the Change or Remove Programs button Proceed with uninstalling by following the instructions that appear on the screen ...

Страница 74: ...nux version for analog I O driver API AIO LNX is supplied as a compressed file linux aio caioXXX tgz on the bundled disk Note XXX represents the driver version Mount the bundled disk as shown below copy the file to an arbitrary directory and decompress the file to install the driver For details on using the driver refer to readme txt and the help file in HTML format extracted by installation To in...

Страница 75: ...ple Programs Sample programs have been prepared for specific basic applications Sample programs for each language are contained in the contec caio samples directory For compiling them refer to the manual for the desired language Uninstalling the driver To uninstall the driver use the uninstall shell script contained in the contec caio directory For details check the contents of the script ...

Страница 76: ... 10µ sec ch Buffer memory 1k Word Conversion start trigger Software external trigger Conversion stop trigger Number of sampling times external trigger software External start signal TTL level Rising or falling edge can be selected by software Digital filter select 1µ sec by software External stop signal TTL level Rising or falling edge can be selected by software Digital filter select 1µ sec by so...

Страница 77: ... Universal key shapes supported 3 Dimension mm 121 69 L x 63 41 H Weight 60g 1 The non linearity error means an error of approximately 0 1 occurs over the maximum range at 0 C and 50 C ambient temperature 2 At the time of the source use of a signal which built in the high speed operational amplifier 3 This board requires 5V power supply from expansion slots it does not operate in the environment o...

Страница 78: ... a circuit block diagram of this board Figure 6 1 Block Diagram 64 single end 32 differential Analog Inputs 4 Digital Inputs 4 Digital Outputs External Trigger Inputs Counter Input Output PCI Bus CN1 DC DC converter FPGA Multiplexer OP Amplifer A D Converter Buffer ...

Страница 79: ...100 nsec Set up time of sampling start Rising edge tSRS 100 nsec Hold time of sampling start Rising edge tHRS 100 nsec Set up time of sampling start Falling edge tSFS 100 nsec Hold time of sampling start Falling edge tHFS 100 nsec Set up time of sampling stop Rising edge tSRP 100 nsec Hold time of sampling stop Rising edge tHRP 100 nsec Set up time of sampling stop Falling edge tSFP 100 nsec Hold ...

Страница 80: ...Figure 6 6 Timing Chart of Counter Output Signal Pulse output Table 6 3 Control Signal Timings Parameter Symbol Time Unit Set up time of counter input Rising edge tSRC 100 nsec Hold time of sampling start Rising edge tHRC 100 nsec Pulse width of counter output signal tPSC 1000 nsec The times listed in Table 6 3 are for standard operating conditions Counter Up Pulse Input tHRC tSRC Counter Compair ...

Страница 81: ...nt and performing the calibration in accordance with the instructions displayed by the calibration program Analog input calibration Analog input calibration requires a reference voltage generator As the analog input has 16 bit resolution use a reference voltage generator with a precision of at least 5 digits after the decimal point Calibrate one channel only for each range that you use Factory set...

Страница 82: ......

Страница 83: ...9 31 Himesato Nishiyodogawa ku Osaka 555 0025 Japan https www contec com No part of this document may be copied or reproduced in any form by any means without prior written consent of CONTEC CO LTD 12152017 10122005 Management No A 51 097 12152017_rev4 Parts No LYFK363 ...