4 Technical Specifications

1.

Display Format:

 Digital tube Display

SpO

2

 Measuring Range:

 0% - 100%

Pulse Rate Measuring Range:

 30 bpm - 250 bpm

Pulse Intensity Display:

 columniation display

2.

Power Requirements:

 2 ×1.5V AAA alkaline battery, adaptable range: 2.6V~3.6V

3.

Power Consumption:

 Smaller than 25 mA

4.

Resolution:

 1% for SpO

2

 and 1 bpm for Pulse Rate

5.

Measurement Accuracy:

 ±2% in stage of 70%-100% SpO

2

, and meaningless when

stage being smaller than 70%. ±2 bpm or ±2% (select larger) for Pulse Rate.

6.

Measurement Performance in Weak Filling Condition:

 SpO

2

 and pulse rate

can be shown correctly when pulse-filling ratio is 0.4%. SpO

2

 error is ±4%,

pulse rate error is ±2 bpm or ±2% (select larger).

7.

Resistance to surrounding light:

 The deviation between the value measured

in the condition of man-made light or indoor natural light and that of darkroom

is less than ±1%.

8. It is equipped with a function switch. The Oximeter can be powered off in case no

finger is in the device.

9.

Optical Sensor

Red light (wavelength is 660nm, 6.65mW)

Infrared (wavelength is 880nm, 6.75mW)

Optical sensor is a light-emitting component, which will affect other 

medical devices appling the same wavelength range.

5 Accessories

• One hanging rope

• Two batteries (optional)

• One User Manual

6 Installation

6.1 View of the Front Panel

Figure 2. Front View 

Figure 3. Batteries Installation

6.2 Battery

Step 1. Refer to Figure 3. and insert the two AAA size batteries properly in the right 

direction.

Step 2. Replace the cover. 

Please take care when you insert the batteries for the improper insertion 

may damage the device. 

6.3 Mounting the Hanging Rope 

Step 1. Put the end of the rope through the hole.

Step 2. Put another end of the rope through the first one and then tighten it.

Figure 4. Mounting the hanging rope 

Figure 5. Put finger in position

7 Operating Guide

7.1  Insert the two batteries properly to the direction, and then replace the cover.

7.2  Open the clip as shown in Figure 5. 

7.3  Let the patient’s finger put into the rubber cushions of the clip (make sure the finger 

is in the right position), and then clip the finger.

7.4  Press the switch button once on front panel.

7.5  Do not shake the finger and keep the patient at ease during the process. 

Meanwhile, human body is not recommended in movement status.

7.6  Get the information directly from screen display.

7.7  In boot-strap state,press button ,and the device is reset.

Fingernails and the luminescent tube should be on the same side.

8 Repairing and Maintenance

• Please change the batteries when the low-voltage displayed on the screen.

• Please clean the surface of the device before using. Wipe the device with medical

alcohol first, and then let it dry in air or clean it by dry clean fabric.

• Using the medical alcohol to disinfect the product after use, prevent from cross

infection for next time use.

• Please take out the batteries if the oximeter is not in use for a long time.

• The best storage environment of the device is - 40ºC to 60ºC ambient temperature

and not higher than 95% relative humidity.

• Users are advised to calibrate the device termly (or according to the calibrating

program of hospital). It also can be performed at the state-appointed agent or just

contact us for calibration.

High-pressure sterilization cannot be used on the device. 

Do not immerse the device in liquid.

It is recommended that the device should be kept in a dry environment. 

Humidity may reduce the useful life of the device, or even damage it.

9 Troubleshooting

10 Key of Symbols

11 Function Specification 

Trouble

Possible Reason

Solution

The SpO

2

 and Pulse Rate 

can not be displayed 

normally.

1. The finger is not

properly positioned.

2. The patient’s SpO

2

 is

too low to be detected. 

1. Place the finger properly

and try again. 

2. Try again; Go to a

hospital for a diagnosis 

if you are sure the device 

works all right.

The SpO

2

 and Pulse Rate 

are not displayed stably.

1. The finger is not placed

inside deep enough.

2. The finger is shaking

or the patient is moving.

1. Place the finger properly

and try again. 

2. Let the patient keep

calm.

The device can not be 

turned on.

1. The batteries are

drained or almost drained. 

2. The batteries are not

inserted properly. 

3. The malfunction of the

device.

1. Change batteries.

2. Reinstall batteries.

3. Please contact the local

service center.

The display is off 

suddenly.

1. The device will power off

automatically when it gets 

no signal within 5 seconds.

2. The batteries are almost

drained.

1. Normal.

2. Change batteries.

Symbol

Description

Type BF

Refer to instruction manual/booklet

 

%SpO

2

The pulse oxygen saturation(%)

    PRbpm

Pulse rate (bpm)

The battery voltage indication is deficient (change the battery 

in time avoiding the inexact measure)
1. No finger inserted

2. An indicator of signal inadequacy
Battery positive electrode
Battery cathode
Power switch

SN

Serial number
Alarm inhibit

WEEE (2002/96/EC)

IP22

Ingress of liquids rank

Display Information

Display Mode

The Pulse Oxygen Saturation (SpO

2

)

Digital

Pulse Rate (BPM)

Digital

Pulse Intensity (bar-graph)

Digital bar-graph display

SpO

2

 Parameter Specification

Measuring range

0%~100%, (the resolution is 1%).

Accuracy

70%~100%: ±2%

Below 70% unspecified

Optical Sensor

Red light (wavelength is 660nm)

Infrared (wavelength is 880nm)

Pulse Parameter Specification

Measuring range

30bpm~250bpm (the resolution is 1 bpm)

Accuracy

±2bpm or ±2% select larger

Pulse Intensity

Range

Continuous bar-graph display, the higher 

display indicate the stronger pulse

Battery Requirement

1.5V (AAA size) alkaline batteries × 2 or rechargeable battery

Battery Useful Life

Two batteries can work continually for 24 hours

Dimensions and Weight

Dimensions

60(L) × 30.5(W) × 32.5(H) mm

Weight

About 50g (with the batteries)

M-35072-GB-Rev

.0.09.17

35072 / CMS50DL

0123

The display SpO

2

 

Low voltage display

Power switch

Pulse rate

Pulse rate
bargraph display

CONTEC MEDICAL SYSTEMS CO., LTD

No.112 Qinhuang West Street, Economic & Technical Development Zone, 

Qinhuangdao, Hebei Province, PEOPLE’S REPUBLIC OF CHINA 

Made in China
Shanghai International Holding Corp. GmbH (Europe)

Eiffestrasse 80, 20537, Hamburg, Germany

Disposal

: The product must not be disposed of along with other domestic waste. 

The users must dispose of this equipment by bringing it to a specific recycling point for 

electric and electronic equipment. For further information on recycling points contact 

the local authorities, the local recycling center or the shop where the product was purchased. 

If the equipment is not disposed of correctly, fines or penalties may be applied in accordance 

with the national legislation and regulations.