4 Technical Specifications
1.
Display Format:
Digital tube Display
SpO
2
Measuring Range:
0% - 100%
Pulse Rate Measuring Range:
30 bpm - 250 bpm
Pulse Intensity Display:
columniation display
2.
Power Requirements:
2 ×1.5V AAA alkaline battery, adaptable range: 2.6V~3.6V
3.
Power Consumption:
Smaller than 25 mA
4.
Resolution:
1% for SpO
2
and 1 bpm for Pulse Rate
5.
Measurement Accuracy:
±2% in stage of 70%-100% SpO
2
, and meaningless when
stage being smaller than 70%. ±2 bpm or ±2% (select larger) for Pulse Rate.
6.
Measurement Performance in Weak Filling Condition:
SpO
2
and pulse rate
can be shown correctly when pulse-filling ratio is 0.4%. SpO
2
error is ±4%,
pulse rate error is ±2 bpm or ±2% (select larger).
7.
Resistance to surrounding light:
The deviation between the value measured
in the condition of man-made light or indoor natural light and that of darkroom
is less than ±1%.
8. It is equipped with a function switch. The Oximeter can be powered off in case no
finger is in the device.
9.
Optical Sensor
Red light (wavelength is 660nm, 6.65mW)
Infrared (wavelength is 880nm, 6.75mW)
Optical sensor is a light-emitting component, which will affect other
medical devices appling the same wavelength range.
5 Accessories
• One hanging rope
• Two batteries (optional)
• One User Manual
6 Installation
6.1 View of the Front Panel
Figure 2. Front View
Figure 3. Batteries Installation
6.2 Battery
Step 1. Refer to Figure 3. and insert the two AAA size batteries properly in the right
direction.
Step 2. Replace the cover.
Please take care when you insert the batteries for the improper insertion
may damage the device.
6.3 Mounting the Hanging Rope
Step 1. Put the end of the rope through the hole.
Step 2. Put another end of the rope through the first one and then tighten it.
Figure 4. Mounting the hanging rope
Figure 5. Put finger in position
7 Operating Guide
7.1 Insert the two batteries properly to the direction, and then replace the cover.
7.2 Open the clip as shown in Figure 5.
7.3 Let the patient’s finger put into the rubber cushions of the clip (make sure the finger
is in the right position), and then clip the finger.
7.4 Press the switch button once on front panel.
7.5 Do not shake the finger and keep the patient at ease during the process.
Meanwhile, human body is not recommended in movement status.
7.6 Get the information directly from screen display.
7.7 In boot-strap state,press button ,and the device is reset.
Fingernails and the luminescent tube should be on the same side.
8 Repairing and Maintenance
• Please change the batteries when the low-voltage displayed on the screen.
• Please clean the surface of the device before using. Wipe the device with medical
alcohol first, and then let it dry in air or clean it by dry clean fabric.
• Using the medical alcohol to disinfect the product after use, prevent from cross
infection for next time use.
• Please take out the batteries if the oximeter is not in use for a long time.
• The best storage environment of the device is - 40ºC to 60ºC ambient temperature
and not higher than 95% relative humidity.
• Users are advised to calibrate the device termly (or according to the calibrating
program of hospital). It also can be performed at the state-appointed agent or just
contact us for calibration.
High-pressure sterilization cannot be used on the device.
Do not immerse the device in liquid.
It is recommended that the device should be kept in a dry environment.
Humidity may reduce the useful life of the device, or even damage it.
9 Troubleshooting
10 Key of Symbols
11 Function Specification
Trouble
Possible Reason
Solution
The SpO
2
and Pulse Rate
can not be displayed
normally.
1. The finger is not
properly positioned.
2. The patient’s SpO
2
is
too low to be detected.
1. Place the finger properly
and try again.
2. Try again; Go to a
hospital for a diagnosis
if you are sure the device
works all right.
The SpO
2
and Pulse Rate
are not displayed stably.
1. The finger is not placed
inside deep enough.
2. The finger is shaking
or the patient is moving.
1. Place the finger properly
and try again.
2. Let the patient keep
calm.
The device can not be
turned on.
1. The batteries are
drained or almost drained.
2. The batteries are not
inserted properly.
3. The malfunction of the
device.
1. Change batteries.
2. Reinstall batteries.
3. Please contact the local
service center.
The display is off
suddenly.
1. The device will power off
automatically when it gets
no signal within 5 seconds.
2. The batteries are almost
drained.
1. Normal.
2. Change batteries.
Symbol
Description
Type BF
Refer to instruction manual/booklet
%SpO
2
The pulse oxygen saturation(%)
PRbpm
Pulse rate (bpm)
The battery voltage indication is deficient (change the battery
in time avoiding the inexact measure)
1. No finger inserted
2. An indicator of signal inadequacy
Battery positive electrode
Battery cathode
Power switch
SN
Serial number
Alarm inhibit
WEEE (2002/96/EC)
IP22
Ingress of liquids rank
Display Information
Display Mode
The Pulse Oxygen Saturation (SpO
2
)
Digital
Pulse Rate (BPM)
Digital
Pulse Intensity (bar-graph)
Digital bar-graph display
SpO
2
Parameter Specification
Measuring range
0%~100%, (the resolution is 1%).
Accuracy
70%~100%: ±2%
Below 70% unspecified
Optical Sensor
Red light (wavelength is 660nm)
Infrared (wavelength is 880nm)
Pulse Parameter Specification
Measuring range
30bpm~250bpm (the resolution is 1 bpm)
Accuracy
±2bpm or ±2% select larger
Pulse Intensity
Range
Continuous bar-graph display, the higher
display indicate the stronger pulse
Battery Requirement
1.5V (AAA size) alkaline batteries × 2 or rechargeable battery
Battery Useful Life
Two batteries can work continually for 24 hours
Dimensions and Weight
Dimensions
60(L) × 30.5(W) × 32.5(H) mm
Weight
About 50g (with the batteries)
M-35072-GB-Rev
.0.09.17
35072 / CMS50DL
0123
The display SpO
2
Low voltage display
Power switch
Pulse rate
Pulse rate
bargraph display
CONTEC MEDICAL SYSTEMS CO., LTD
No.112 Qinhuang West Street, Economic & Technical Development Zone,
Qinhuangdao, Hebei Province, PEOPLE’S REPUBLIC OF CHINA
Made in China
Shanghai International Holding Corp. GmbH (Europe)
Eiffestrasse 80, 20537, Hamburg, Germany
Disposal
: The product must not be disposed of along with other domestic waste.
The users must dispose of this equipment by bringing it to a specific recycling point for
electric and electronic equipment. For further information on recycling points contact
the local authorities, the local recycling center or the shop where the product was purchased.
If the equipment is not disposed of correctly, fines or penalties may be applied in accordance
with the national legislation and regulations.