CONTEC MEDICAL SYSTEMS CONTEC08E Скачать руководство пользователя страница 2

Do not move during measurement, it will have a delayed effect on the patient’s blood flow. 

The device need to be placed for 2 hours from the minimum storage temperature to being ready for its intended use. 

The device need to be placed for 4 hours from the highest storage temperature to being ready for its intended use. 

  Note 

 

The following conditions may also cause changes in the blood pressure measurement value.

 

Take the measurement in one hour after meal or after drinking alcohol, coffee or after smoking, exercise, bathing; 

Using incorrect posture such as standing or lying down, etc; 

The patient speak or move his body during measurement; 

When measuring, the patient is nervous, excited, emotional instability; 

The room temperature rise or fall sharply, or the environment of measurement often changes; 

Measuring in a moving vehicle; 

The high and low location of cuff will cause changes in measurement results; 

Continuous measurement for a long time. 

7.2 Applying the Cuff 

Both left and right arm can be measured. 

Bare your arm or cloth close-fitting clothing during measurement. 

Carry out the operation in a room with comfortable temperature. 

When measuring, take the thick clothes off instead of rolling up the sleeves.

 

In order to measure accurately, pay attention to applying the cuff properly (left arm). 

 Insert the arm cuff air plug in the cuff socket of sphygmomanometer. 

 Stretch cuff into a barrel for the arm can conformable enter into the barrel 

 Left arm penetrate through the cuff, the air tube of the cuff will pass the top of your palm. 

 Wrap the cuff to your upper arm. Make the air tube inside the forearm and aligned with your middle finger. 

 The bottom of the cuff should be approximately 2cm~3cm above your elbow. 

 Be fixed with cloths, and wrapped tight cuff, the arm and the cuff should not have gaps. 

       

       

 

                                                         

                                                             

 

         

 

                                                         

                                                               

 

7.3 Measurement BP 

Under “OFF” state, press “START/STOP” button to start measuring. 

During measurement, please keep correct pose and quiet state, the body could not move.The “Movement” icon appears 

if patient moves, and continue measuring may lead to inaccurate measurement. 

If you want to abort the measurement   

Press

START/STOP

button, the device will stop inflating, and release the air from the cuff. 

Display  the  measurement  results  after  finishing  measuring.

The  pressure  bar  on  the  right  side  visually  demonstrates  the 

pressure level.

 

7.4 Confirm the Measurement Value 

The World Health Organization has established globally accepted standards for the assessment of hypertension readings.(In the 

clinic environment) 

Blood pressure level 

Systolic pressure 

Diastolic pressure 

Normal 

Pressure

130mmHg 

Pressure

85mmHg 

Normal Systolic Value 

130mmHg

Pressure

139mmHg 

85mmHg

Pressure

89mmHg 

Mild Hypertension 

140mmHg

Pressure

159mmHg 

90mmHg

Pressure

99mmHg 

Moderate Hypertension 

160mmHg

Pressure

179mmHg 

100mmHg

Pressure

109mmHg 

Severe 

Hypertension/High 

Blood 

Pressure 

180mmHg

Pressure 

110mmHg

Pressure 

Pressure bar at the right side 

The SYS higher than 135mmHg or the DIA higher than 85mmHg are used as the criteria of hypertension(In the home environment), 

and the pressure bar at the right side lights up in red. 

The SYS lower than 135mmHg and the DIA lower than 85mmHg are used as the criteria of normal pressure, and the pressure bar 

at the right side lights up all in green. 

The number of lights represents the blood pressure range. 

*Self-diagnosis and treatment using measured results may be dangerous. Follow the instructions of your physician. 

7.5 Upload date(optional for devices with Bluetooth function) 

The data stored can be uploaded to master device by Bluetooth 

Time  sync  icon  lights  up  if  the  device  time  is  not  synced,  and  the  measured  data  cannot  be  uploaded  to  the  terminal 

equipment  via the Bluetooth. After syncing the time by Bluetooth connection  of the  device,  Time sync  icon  goes  off, 

and the measured data can be uploaded 

After uploading the data to the master device, the local data will be deleted. 

  Note 

 

Wait at least 4-5 minutes between measurements.

 

 

When repeatedly measuring, the accurate blood pressure value may not be measured due to congestion in the arm. 

Please measure after the blood flow is smooth. 

 

When the screen displays Err, the measure can't be carried out correctly. 

 

Irregular  pulse  icon  is  displayed  in  the  measurement  results  if  the  pulse  internal  is  irregular  during  measuring, 
which  may  cause  it  is  unable  to  take  measurement  correctly.  Please  keep  quiet  and  remeasure.  If  the  irregular 

pulse icon appears frequently, please consult a doctor. 

 

The  minimum  value  of  the  patient’s  physiological  signal  is  the  minimum  limit  that  the  device  can  measure. The 

device may obtain inaccurate measurement results when it is operated below the minimum amplitude or minimum 
value of the patient’s physiological signal. 

*The  device  will  automatically  turn  off  after  five  minutes  in  which  there  is  no  operation  to  the  device,  even  if  you 
forget to turn the power off. 

Chapter8 Memory Function 

The device can store NIBP values automatically, display up to 199 set of measurement results. 

If 199 set of measurement data have been stored in current device, when saving the 200th set of data, the earliest set of 

data will be overwritten. If no measurement values, the memory values can be not numerated. 

Memory function can not be used during measuring. 
When there is no measurement values, “

---

” will display on the review interface. 

8.1 Review the Memory Value 

1.Under “OFF” state, press “M” button to display the average value of the latest three set of data, when the number of 
measurement data is less than three groups, it will supplement automatically. Continue to press “M” button in current 

interface to view all measurement records. 

8.2 Delete Memory Values 

1.Users can delete all memory values of the current user instead of separately delete one memory value 
2.Under the  memory interface, press “M” button and “START/STOP” button simultaneously for more than 5 s, after 
“DEL” appears on the screen, all memory values will be deleted. 

  Caution 

 

When querying the measurement records, please press “M” button continuously to query one by one. 

Chapter9 Key and Symbols 

Your device may not contain all the following symbols. 

Signal 

Description 

Signal 

Description 

 

Attention! Please refer to the accompanying 
document (the user manual). 

 

Attention! Please refer to the accompanying 
document (the user manual). 

SYS 

Systolic pressure 

DIA 

Diastolic pressure 

MAP 

Mean blood pressure 

PUL 

Pulse rate (bpm) 

IP20 

Enclosure protection grade 

EMC 

Electromagnetic compatibility 

 

  Recyclable 

P/N 

Material code of manufacturer 

 

Batch code 

 

Use by date 

 

This way up 

 

Fragile, handle with care 

 

Keep dry 

 

Storage atmospheric pressure limitation 

 

Storage temperature limitation 

 

Storage humidity limitation 

 

 

Manufacturer 

 

Date of manufacture 

 

Batteries Power   

SN 

Serial number     

 

Flating 

 

Deflating 

 

Waste disposal mark, this symbol indicates 
that the waste of electrical and electronic 
equipment can not be disposed as an 
unclassified municipal waste and must be 
recovered separately. 

 

This item is compliant with Medical Device 
Directive 93/42/EEC of June 14, 1993,a directive 
of the European Economic Community. 

 

Class II equipment 

 

Type BF applied parts 

  European Representative 

 

Irregular pulse 

 

Socket for power adapter 

 

Interface for connecting cuff 

       

 

Voice closed 

 

Voice enabled 

 

Large movement during measurement 

 

Cuff tied properly 

 

Artery indicator label 

 

Bluetooth 

 

time sync icon 

 

 

 

Chapter10 Error Message   

When the high pressure position appears "Err" and the low pressure position appears the error number, the measurement 

is not normal. 

Error Mark 

Causes   

Solutions 

Err2 

Err15 

Function abnormal 

Please contact us 

Err4 

Low battery 

Please replace the battery or link adapter 

Err6 

The cuff is not wrapped correctly. 

Wrap the cuff correctly (refer to Chapter 7) 

Err7 

Cuff leakage 

Replace with a new cuff 

Err8 

Air pressure error 

Keep arm, body still, measure again 

Err9 

The pulse signal is too weak or the cuff is loose. 

Wrap the cuff correctly (refer to Chapter 7) 

Err10 

Out of measure range 

Keep arm, body still, measure again 

Err12 

Cuff is blocked or squeezed 

Wrap the cuff correctly (refer to Chapter 7) 

Err11 

Err13 

The signal amplitude is too big owing to the arm 

or body moving or other reasons when measuring 

Keep arm, body still, measure again 

Err16 

Err19 

It takes too much time 

 

Chapter11 Troubleshooting 

 

Abnormal 

Phenomenons 

Causes 

Solutions 

BP measurement 
values too high or 
too low. 

Cuff is not connected correctly. 

Correctly connect cuff.   

Talk or move arm in measurement 

Keep quiet and restart a measurement. 

The turnup close oppress the arm 

Take off the clothes, and restart a measurement 

No pressure 

Cuff leakage 

Buy a new cuff. 

The cuff windpipe is not correctly connected 
with cuff 

Correctly connect. 

Cuff not inflate 

Contact us. 

Cuff deflate in short 
time 

Loose cuff 

Correctly tangle cuff. 

It can not carry on measurement ,even if press the measurement 
button 

Return on the power and restart a 
measurement. 

Abruptly turn the 
power off in adding 
pressure 

No use for a long time, the batteries can be 
exhausted owing to the changed temperature 

Replace all four batteries with new ones. 

Hold the on/off 
button but can not 
start the device 

Batteries are worn 

Replace all four batteries with new ones. 

The battery polarities is reversed   

Check the battery installation for proper 
placement of the battery polarities. 

Cuff inflation start before press the measurement button 

Stop using the device and contact us. 

Cuff never deflation 

Stop using the device and contact us. 

Air pressure error 

Deflation error 

Pull out the cuff to deflate. Stop using the 
device and contact us. 

Others   

Keep arm, body still, measure again. 

No press value displayed or the value unaltered when cuff inflating 

Pull out the cuff to deflate. Stop using the 
device and contact us. 

Other phenomenon 

Switch on the power once again and restart an 
operation. 
Replace the batteries. 
If no, please contact us. 

 

Chapter12 Maintenance, Cleaning and Keeping 

*Please do obey the precautions and correct operating methods in this user manual. Otherwise, we will not responsible for 

any fault. 

  Warning 

 

Remove the batteries before cleaning. The accessories and main unit must be separated for cleaning. 
Maintenance is not allowed during device using. 

Do not squeeze the rubber tube on the cuff. 

  Caution 

 

 

High pressure disinfection to the device and accessories is not allowed. 

 

Do not let water or cleaning agent flow into the socket to avoid device damage. 

 

Do not soak the device and accessories in liquid. 

 

If any damage or deterioration of the device and accessories is found, please do not use it. 

Maintenance:

 

 

Clean the  device and accessories regularly. It is recommended to clean them  every one  month. When the  device  or 

accessory  gets dirty, use a dry and soft cloth to  wipe.  If they are  very  dirty, it is available to  dip the soft cloth into 
water or mild detergent, and wring out, then use the cloth for cleaning. 

 

The  device  shall  be  inspected  and  calibrated  regularly  (or  according  to  inspection  standard  of  hospital).  The 
inspection  can  be  carried  out  in  appointed  institutions,  or  by  professional  personnel  or  contact  us  for  inspection. 
Under the setting interface, Press the “START/STOP” button  once, after “CAL” appears on the screen, press  “M” 
button for more than 15 s to enter the static pressure interface. 

  Advice 

   

 

Do not use gasoline, volatile oil, diluent, etc. to wipe the device. 

 

Do not clean or wet the cuff. 

Storage: 

  Advice 

   

 

Do not expose the device in direct sunlight for long time, otherwise the display screen maybe damaged. 

 

The basic performance and safety of the device are not affected by the dust or cotton wool in home   

environment,while the device shall not be placed where with high temperature, humidity or dusty. 

 

Aged cuff may result in inaccurate measurement, please replace the cuff periodically according to the user manual. 

 

To avoid device damage, keep the device out the reach of children and pets. 

 

Avoid the device close to extreme high temperature such as fireplace, otherwise the device performance may be affected. 

 

Do not store the device with chemical medicine or corrosive gas. 

 

Do not place the device where there is water. 

 

Do not place the device where with slope, vibration or impact 

 

Take the batteries out if the device is not to be used for three months or longer. 

Chapter13 NIBP Specification 

 

Name 

Electronic Sphygmomanometer

 

The degree of protection 
against ingress of water 

IP20 

Display 

LED 

NIBP Specifications 

Measurement Method 

Oscillometric method 

Working mode 

Automatic 

Operation mode 

Continuous operation 

Pressure Range 

0~297 mmHg(0~39.6 kPa) 

Measurement range 

Pressure 

SYS

30

270 mmHg(4

36 kPa) 

DIA

10

220 mmHg(1.3

29.3 kPa) 

Pulse: 40~240/min 

Inflation 

160±5 mmHg(21.33±0.67 kPa) 

Overpressure protect 

297±3 mmHg(39.6±0.4 kPa) 

Resolution 

Pressure: 1 mmHg(0.133 kPa) 

Accuracy 

Static pressure: ±3 mmHg(±0.4 kPa) 

Error 

The BP value measured by the device is equivalent with the measurement value of 
Stethoscopy, perform clinical verification in accordance with the requirements in ISO 
81060-2: 2013, whose error meets the followings: 
Maximum mean error: ±5 mmHg 
Maximum Standard deviation: 8 mmHg 

Operating Temperature/ 
Humidity 

  +5º

C~40 º

C            .    15%RH~85%RH(no condensation) 

Transport 

Transport by general vehicle or according to the order contract, avoid pounded, shake and 
splash by rain and snow in transportation. 

Storage 

Temperature: -20 º

C~+55 º

C; Relative humidity: ≤95 %((no condensation)); No corrosive 

gas and drafty. 

Atmospheric pressure 

700 hPa~1060 hPa 

Power supply 

4 "AA" alkaline batteries, AC Adapter(AC, 100 V-240 V, optional) 

Rated current 

≤ 600 mA 

Battery life 

When  the  temperature  is  23  º

C,  limb  circumference  is  270  mm,  the  measured  blood 

pressure is normal, 4 "AA" alkaline batteries cab be used about 300 times. 

Main Unit Dimensions 

129*101*72 mm   

 

Main Unit Weight 

300 gram(without batteries) 

Safety classification 

Class 

 equipment (power supplied by power adapter)/Internally powered equipment 

(power supplied by batteries) 
Type BF applied part 

Service life 

The service life of the device is five years or 10000 times of BP measurement. 

Date of manufacturer 

See the label 

Accessories 

 

Standard Configure: 
Adult Cuff: limb circumference 22-32 cm (upper arm center)   
User Manual, four "AA" alkaline batteries 
Optional Configure

 

AC Adapter: 
Input: voltage: AC 100 V~240 V    frequency: 50 Hz/60 H      Rated current: AC 150 mA 
Output: DC 5.0 V±0.2 V      1.0 A 
Extra large adult Cuff: the range of limb circumference is 18-26 cm (middle part of upper 
arm) 
the range of limb circumference is 22-30 cm (middle part of upper arm) 
the range of limb circumference is 22-43 cm (middle part of upper arm) 
the range of limb circumference is 32-43 cm (middle part of upper arm) 

 

Appendix 

Table 1:   

Guidance and manufacturer’s declaration –electromagnetic emission 

The device is intended for use in the electromagnetic environment specified below. The purchaser or the user of the 
device should assure that it is used in such environment. 

Emission test 

Compliance 

RF emissions CISPR 11 

Group 1 

RF emissions CISPR 11 

Class B 

Harmonic emissions 
IEC 61000-3-2 

Class A 

Voltage fluctuations/flicker emissions 
IEC 61000-3-3 

Applicable 

Table 2: 

Guidance and manufacturer’s declaration-electromagnetic immunity 

The device is intended for use in the electromagnetic environment specified below. The purchaser or the user of the 
device should assure that it is used in such environment. 

Immunity test 

IEC60601 test level 

Compliance level 

Electrostatic 

discharge 

(ESD)   
IEC 61000-4-2 

±8kV contact 
± 15 kV air 

±8kV contact 
±15kV air 

Electrical fast 
transient/burst 
IEC 61000-4-4 

±2kV for power supply lines 
± 1 kV for input/output line 

±2kV for power supply lines 
Not Applicable 

Surge   
IEC 61000-4-5 

±1 kV lines to lines 
±2 kV lines to earth 

±1 kV lines to lines 
Not Applicable 

Voltage dips, short 
interruptions and voltage 
vatiations on power supply 
input lines 
IEC 61000-4-11 

<5%UT(>95%dip in UT) for 0.5 cycle   
40% UT(60%dip in UT) for 5 cycle   
70%UT(30%dip in UT) for 25 cycle   
<5%UT(>95%dip in UT) for 5 sec 

<5%UT(>95%dip in UT) for 0.5 cycle   
40% UT(60%dip in UT) for 5 cycle   
70%UT(30%dip in UT) for 25 cycle   
<5%UT(>95%dip in UT) for 5 sec 

Power  frequency  (50  / 
60Hz) magnetic field   
IEC 61000-4-8 

30 A/m 

30A/m 

Table 3:   

Guidance and manufacturer’s declaration – electromagnetic immunity 

The  device  is  intended  for  use  in  the  electromagnetic  environment  specified  below.  The  customer  the  user  of  the 

device should assure that it is used in such environment. 

Immunity test 

IEC 60601 test level 

Compliance level 

Conducted RF 
IEC61000-4-6 

3 V 

0,15 MHz – 80 MHz 
6 V in ISM bands between 

0,15 MHz and 80 MHz 

3 V 

0,15 MHz – 80 MHz 
6 V in ISM bands between 

0,15 MHz and 80 MHz 

Radiated RF 

IEC61000-4-3 

10 V/m 80 MHz- 2.7 GHz 

10 V/m80 MHz- 2.7 GHz 

NOTE 1  At 80 MHz and 800 MHz, the higher frequency range applies. 
NOTE 2  These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and 

reflection from structures, objects and people. 

a    Field  strengths  from  fixed  transmitters,  such  as  base  stations  for  radio  (cellular/cordless)  telephones  and  land 

mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with 
accuracy.  To  assess  the  electromagnetic  environment  due  to  fixed  RF  transmitters,  an  electromagnetic  site  survey 

should  be  considered.  If  the  measured  field  strength  in  the  location  in  which  the  the  device  is  used  exceeds  the 
applicable  RF  compliance  level  above,  the  the  device  should  be  observed  to  verify  normal  operation.  If  abnormal 

performance is observed, additional measures may be necessary, such as reorienting or relocating the the device. 

Table 4

 

Guidance and manufacturer’s declaration - electromagnetic Immunity 

The device is intended for use in the electromagnetic environment specified below. The customer or the user of the 
device should assure that it is used in such an environment 

Radiated RF 
IEC61000-4-3 
(Test 
specifications 
for 
ENCLOSUR
EPORT 
IMMUNITY 
to RF wireless 
communicatio
ns equipment) 

Test 
Frequency 
(MHz) 

Band a) 
(MHz) 

Service a) 

Modulat
ion b) 

Modulat
ion b) 
(W) 

Distan
ce 
(m) 

IMMUNIT

TEST 
LEVEL 
(V/m) 

385 

380– 
390 

TETRA 400 

Pulse 
modulat
ion b) 
18 Hz 

1,8 

0,3 

27 

450 

380– 
390 

GMRS 460, 
FRS 460 

FM c) 
±  5  kHz 
deviatio

kHz 

sine 

0,3 

28 

710 

704– 
787 

LTE Band 13, 
17 

Pulse 
modulat
ion b) 
217 Hz 

0,2 

0,3 

745 

780 

810 

800– 
960 

GSM 800/900, 
TETRA 800, 
iDEN 820, 
CDMA 850, 
LTE Band 5 

Pulse 
modulat
ion b) 
18 Hz 

0,3 

28 

870 

930 

 

1720 

1700–199

GSM 1800; 
CDMA 1900; 
GSM 1900; 
DECT; 
LTE Band 1,3,4,25; 
UMTS 

Pulse 
modulat
ion b) 
217 Hz 

0,3 

28 

1845 

1970 

2450 

2400–257

Bluetooth,WLAN,802.11 
b/g/n,RFID 

2450,LTE 

Band 7 

Pulse 
modulat
ion b) 
217 Hz 

0,3 

28 

5240 

5100–580

WLAN 802.11a/n 

Pulse 
modulat
ion b) 
217 Hz 

0,2 

0,3 

5500 

5785 

NOTE If necessary to achieve the IMMUNITY TEST LEVEL, the distance between the transmitting antenna and the 

ME EQUIPMENT or ME SYSTEM may be reduced to 1 m. The 1 m test distance is permitted by IEC 61000-4-3. 

a) For some services, only the uplink frequencies are included. 
b) The carrier shall be modulated using a 50 % duty cycle square wave signal. 
c) As an alternative to FM modulation, 50 % pulse modulation at 18 Hz may be used because while it does not 
represent actual modulation, it would be worst case. 

The MANUFACTURER should consider reducing the minimum separation distance, based on 
RISK MANAGEMENT, and using  higher IMMUNITY TEST LEVELS that are appropriate for the reduced  minimum 
separation distance. Minimum separation distances for higher IMMUNITY TEST LEVELS shall be calculated using the 
following equation:           

                                                                                     

P

d

6

E

 

Where  P  is  the  maximum  power  in  W,  d  is  the  minimum  separation  distance  in  m,  and  E  is  the  IMMUNITY  TEST 
LEVEL in V/m. 