Chapter 14
ERROR MESSAGE
Error message will be displayed in the screen if there is something wrong when measuring. The causes and solutions are
shown as follows:
Error Message
Causes
Solutions
Self-test failure
System failure
Function abnormal
Please contact us
Loose cuff
Cuff is not connected correctly.
Correctly connect cuff (refer to Chapter 9)
Air leakage
Cuff plug fall off
Make sure the cuff plug is securely inserted
in the windpipe (refer to Chapter 9)
Air pressure error
Air pressure error
Refer to the troubleshooting
Weak signal
The pulse signal is too weak or the cuff
is loose.
Correctly connect cuff (refer to Chapter 9)
Overpressure
Cuff is blocked or squeezed
Correctly connect cuff (refer to Chapter 9)
Excessive movement
Over range
Saturated signal
The signal extent is too big owing to the
limbs or body moving or other reasons
when measuring
Keep limbs, body still, measure again
Time out
It takes too much time