C931 • C932 • C933
10 / 47
(C932, C933)
Dissolved oxygen measurement
1.
Select [O
2
] by pressing
MODE
. The display will immediately show the measured
value according to the previous calibration. Should you want to recalibrate,
press
CAL
.
2.
The display shows the salinity correction. Select the proper value and press
CAL
. Leave salinity correction to zero unless you are going to measure in
heavily salted solutions such as e.g. sea-water (35 g/l). Select [CALIBRATE],
press
CAL
and follow the instructions on the screen until the calibration is
finished.
3.
The electrode exposed to the air reaches an equilibrium corresponding to the
partial pressure of oxygen and thus to saturation in water at the given temper-
ature. The instrument shows the measured saturation, current, temperature,
and will calibrate automatically when readings are stable.
4.
After rinsing the electrodes with distilled water, immerse them in the samples
and read the display. Stirring the solution during the measurements promotes
the homogeneity and is
obligatory
! The advection rate must be at least 10
cm/s.
5.
Rinse the electrode always after use and store it in distilled water.
• A blinking decimal point warns you for unstable measurements. Wait to read
the display!
• Stirring the solution during the measurements promotes the homogeneity and
is therefore always recommended.
• Press
Ú
to change the resolution from 0.1 to 0.01 mg/l or from 1 to 0.1 %.
(C932, C933)
%O
2
measurement
1.
Select [%O
2
] by pressing
MODE
. The display will immediately show the meas-
ured value according to the previous calibration. Should you want to recali-
brate, press
CAL
. Proceed as for mg/l.
(C932, C933)
Air pressure measurement
1.
Select [hPa] by pressing
MODE
. Should you want to recalibrate, press
CAL
.
2
. Adjust to the real air pressure and press
CAL
.
(C932, C933)
Principle
The oxygenmeter and its electrode function according to the Clark principle with
silver as cathode and zinc as anode in an electrolyte cell. Oxygen gas present
in the electrolyte is reduced to OH ions at the cathode. The resulting current is
diffusion limited and therefore proportional to the oxygen concentration in the
sample solution. This current is amplified, corrected, and displayed in mg/l, ppm
or % dissolved oxygen.
(C932, C933)
Interferences
All substances which can diffuse through the membrane and for which 800 mV
potential suffices for polarographic reduction, will be reduced in the electrode.
This will give a corresponding current contribution, if they are present.
Interference can be caused by ions entering the electrode through porous or
mechanically damaged membranes and by diffusion of other reactive gases apart
from oxygen, e.g. CO
2
, Cl
2
, SO
2
, and H
2
S. These substances react in undesired
manner with the electrode. Acidic or basic gases change the pH value of the
electrolyte solution and thus disturb the reading, particularly when measuring
small oxygen concentrations. High salt concentrations in the sample solution can
falsify readings too.
Содержание C931
Страница 1: ...MANUAL HANDLEIDING MODE D EMPLOI ANLEITUNG C931 C932 C933 July 2003 ...
Страница 2: ...C931 C932 C933 2 47 ...