21
Each of the five voltage ranges presents an input impedance of 10 MOhm
(in parallel with < 100 pF). AC-voltage input is AC-coupled.
As soon as “ - “ is set before the display value during DC voltage measu-
rement, the voltage is negative (or the test leads are confused).
Hint!
During AC voltage measurement the right sub-display will show the dB-
value of the AC voltage, the middle sub-display the frequency and the
main display the present measuring value.
Due to the sensitive nature of the input terminal your meter might dis-
play any small values (not connected with the object). This residual value
will not affect the actual measurements. But Auto-Power-Off is not
active when the “phantom-measurements“ are displayed (for AC volta-
ge „mV“ and „V“).
During DC voltage measurement the three sub-displays below the main
display are active.
In the left sub-display the measuring value appears 1 second later, in the
middle sub-display the measuring value appears 2 seconds later and in
the right sub-display the measuring value appears 3 seconds later. This
means the left sub-display shows the measurement value which was 1 second
earlier on the main display, the middle sub-display shows the reading
which was 2 seconds earlier in the main display and the right sub-display
shows the value which was 3 seconds earlier on the main-display.
~
V
Voltage Source
Object
20
C
= Centigrade Degree
F
= Fahrenheit Degree
µF
= Microfarad (exp.-6)
nF
= Nanofarad (exp.-9)
KOhm = Kiloohm (exp.3)
MOhm = Megaohm (exp.6)
dB
= Decibels
5. Measurement
5.1 Voltage Measurement
Attention!
Never exceed the maximum input limits. of max. 1000 VDC or 750 VACrms.
Do not touch circuits or parts of circuits, if you measure voltage more
than 25 VACrms or 35 VDC.
To measure DC or AC voltage proceed as follows
1. Set the rotary function switch to the desired position (mV or V)
2. Press DC Ohm/AC (
) key, depending if you want to measure DC or
AC voltage. As soon as “AC“ appears in the display, you are in the AC
voltage range.
3. Connect the test-lead tips with the object to be measured (load, cir-
cuit etc).