2-6

AER DEFENSE

TM

S M O K E E V A C U A T O R

9.  Activate the unit by:

a.  Depressing and releasing the footswitch 

(if connected), select Continuous 

Activation Button on front panel or 

b.  Activating CUT, COAG or BIPOLAR 

on the ESU.  

NOTE:  Do not activate before both the 

AER DEFENSE™ and ESU have been 

turned on for at least 30 seconds.

10. Adjust the suction level to the desired setting 

by pressing the up or down arrow buttons on 

the control panel while the unit is activated.   

Noise created by the smoke evacuator may 

be minimized by selecting the lowest suction 

setting that effectively clears the sterile field of 

surgical smoke. 

11. Deactivate the unit by: 

a.  Depressing and releasing the footswitch 

(if connected), selecting Continuous 

Activation button or 

b. Deactivating the CUT, COAG or 

BIPOLAR function on the CONMED 

ESU.

12. Replace the CONMED FilterOne™ when the 

ORANGE LED is illuminated in the Filter 

Life Indicator. Failure to change the filter 

will affect the performance and efficacy of the 

system. 

Содержание AER DEFENSE

Страница 1: ...Operator s Manual C M ON ED E L E C T R O S U R G E R Y TM AER DEFENSETM S M O K E E V A C U A T O R...

Страница 2: ...tica NY 13502 5994 USA Patent Pending For Technical Service or Return Authorization Phone 303 699 7600 1 800 552 0138 Extension 5274 Fax 303 699 1628 For Customer Service or to order parts phone 1 800...

Страница 3: ...1 5 1 5 2 Rear Panel 1 6 2 0 Operation 2 1 2 1 System Controls 2 1 2 1 1 Control Panel 2 2 2 1 2 Rear Panel 2 3 2 2 FilterOne Installation 2 4 2 3 Set up and Operation 2 5 2 3 1 AER DEFENSE ESU Activa...

Страница 4: ...AER DEFENSETM S M O K E E V A C U A T O R This page intentionally left blank...

Страница 5: ...es toxic organic gases rather than water vapor and provides optimal odor removal The third stage of filtration is a ULPA Ultra Low Penetration Air grade filter whose high tech design captures particul...

Страница 6: ...ded receptacle of the same voltage as listed on the nameplate on the smoke evacuator WARNING To avoid risk of electric shock this equipment must only be connected to a supply mains with protective ear...

Страница 7: ...will void the warranty If filter is inserted into the AER DEFENSE and the filter is not recognize and no green LED lights illuminate check to make sure filter is inserted properly and completely If fi...

Страница 8: ...s device to ensure compatibility with other electrical devices List of compatible CONMED accessories 60 8054 001 Adapter non sterile 60 6828 005 ClearVac 60 6829 005 ClearVac 60 6828 001 ClearVac 60 6...

Страница 9: ...ter 1 5 Explanation of Symbols 1 5 1 Control Panel 1 4 1 Accessory Diagram All adapters tubing ClearVac and GoldVac smoke pencils attach to the front nozzle on the FilterOne Image Shown 60 7510 005 Go...

Страница 10: ...icing to qualified service personnel Refer to Instruction Manual Booklet for critical safety instruction The color of the symbol is blue 9 Defibrillation Proof Type CF Applied part 9083 Unit functions...

Страница 11: ...s default mode when none of the available specialty modes have been selected and illuminated on the control panel Laparoscopic Continuous Automatic or Footswitch While in Normal Mode the AER DEFENSE S...

Страница 12: ...SU Automatic Mode Laparoscopic Mode indication of the status of the life of the filter in use Each LED segment represents 1 8 or 12 5 of the filter life The life of the filter is a minimum of 10 hours...

Страница 13: ...e 2 1 2 Rear Panel T10 0A H 250V REPLACE FUSES AS MARKED 100 120V 50 60Hz 12A 120V 6 3A ESU Activation Outlet RS232 Connector AER DEFENSE Fuses AER DEFENSE Mains Power Inlet Pneumatic Footswitch Conne...

Страница 14: ...e the unit If the AER DEFENSE is turned back on the Orange LED in the filter life indicator will flash and there is an audible warning 5 beeps to indicate that a new filter must be installed The unit...

Страница 15: ...ange this filter may result in decreased efficiency and contamination of the electric motor vacuum pump and sound absorbing media within the unit 2 3 Set up and Operation with Instant ESU Activation T...

Страница 16: ...or down arrow buttons on the control panel while the unit is activated Noise created by the smoke evacuator may be minimized by selecting the lowest suction setting that effectively clears the sterile...

Страница 17: ...O K E E V A C U A T O R Figure 2 3 1 AER DEFENSE ESU Activation Connection Set Up CORD TO ESU ESU ACTIVATION AER DEFENSE Power Cord ESU Activation Cord T10 0A H 250V REPLACE FUSES AS MARKED 100 120V...

Страница 18: ...2 8 AER DEFENSETM S M O K E E V A C U A T O R AER DEFENSE Power Cord ESU Activation Cord...

Страница 19: ...th 8 2 20 8 cm Length 18 7 47 5 cm Weight 20 lbs 9 07 kg Filter ULPA grade with Carbon Filtration Efficiency 99 999 0 12 m particles Tubing Sizes 7 8 22 mm 6 4 mm Operating Temperature 10 C to 30 C Tr...

Страница 20: ...2 10 AER DEFENSETM S M O K E E V A C U A T O R This page intentionally left blank...

Страница 21: ...Troubleshooting the AER DEFENSE Problem Potential Cause s Corrective Actions Smoke Evacuation System is on but suction is minimal 1 FilterOne not seated completely 2 FilterOne clogged 3 Vacuum hose cl...

Страница 22: ...3 2 AER DEFENSETM S M O K E E V A C U A T O R This page intentionally left blank...

Страница 23: ...cover of this manual Note the R A number on the outside of the carton 4 2 Service Ordering Parts To reorder obtain replacement parts or to return a unit for service call Customer Service at the number...

Страница 24: ...4 2 AER DEFENSETM S M O K E E V A C U A T O R This page intentionally left blank...

Отзывы: