![ConLAB ACT-2I Скачать руководство пользователя страница 2](http://html1.mh-extra.com/html/conlab/act-2i/act-2i_operators-manual_2652542002.webp)
2.
CONNECTION DIAGRAM
ACT-2V
ACT-2I
3.
DIMENSIONS
4. Specifications
ACT-2I
INPUT:
AC current
INPUT SPAN:
0-1Aac, 0-5Aac (RMS)
INPUT BURDEN:
0.26VA at 5Aac
INRUSH CURRENT:
40Aac (RMS) max. for 8 sec/10min
75Aac (RMS) max for 1 sec/ 10min
ISOLATION:
2.5KV (RMS)/1 minute
ACT-2V
INPUT:
Voltage input
INPUT SPAN:
0-400Vac (RMS)
INPUT BURDEN:
1.4VA at 250Vac, 0.3VA at 120 Vac
ISOLATION:
2.5KV (RMS)/1 minute
GENERAL SPECIFICATIONS
OUTPUT RANGE:
4-20mA
TEST TERMINALS:
40-200mV represents 4-20mA
SUPPLY VOLTAGE:
10-32Vdc, reverse polarity protected
SUPPLY AND LOAD VARIATION EFFECT:
< ±0.03% of span for full change
LOOP DRIVE:
Rmax (
Ω
) = (Vsup-10)/0.02
RESPONSE TIME:
220msec. (0-98% of span)
OUTPUT RIPPLE:
Better than 0.1% of span RMS
FREQUENCY:
40- 60 and 400 Hz
ACCURACY (2-100% of input span):
±0.1% of span for pure sinusoidal input
FINE ADJUSTABILITY:
Zero:
±35% of span (minimum)
Span:
±25% of span (minimum)
CMR:
122 dB typical, 117 dB
minimum
TEMPERATURE STABILITY:
±0.015% of span/1
°
C
OPERATING AMBIENT:
32 to 158
°
F
(0-70
°
C)
STORAGE TEMPERATURE:
-13 to 185
°
F
(-25 to +85
°
C)
HUMIDITY:
5-95% relative humidity, non-condensing
PROTECTION LEVEL:
Box:
According to IP-50 DIN 40050
Terminals: According to IP-10 DIN 40050
MOUNTING:
Standard 35mm DIN rail
WEIGHT:
0.5lb (0.2Kg)