TROUBLESHOOTING

Sympton

Cause

Unit  is  plugged  in  but  no  lights  are  lighting  on  the 
Alarm

1.  Check to see that there is power at the source.
2.  Check that the power switch on the unit in ON.
3.  Check to see that the internal fuse is OK.
4.  If the Alarm has been re-configured from the 

factory settings, check that the DIP switches were 
properly set.  Refer to "Configuring the Remote 
Alarm" section of the manual.

Audible  Alarm  begins  to  sound  as  soon  as  power  is 
turned ON.

1.  Check that the Input Connector is plugged in 

properly.

2.  Check that the connector to the Autoswitch is 

plugged in.

3.  Check that the cable between the Autoswitch and 

Alarm is not cut or pinched somewhere.

4.  Check to see that all the wires in the cable are 

properly soldered to the connectors.

5.  Check to see how many status indicator lights are 

lighted.  Compare this to how many systems are 
wired to the Alarm.  They must be the same.  It 
may be necessary to reconfigure the DIP switches.  
See "Configuring the Remote Alarm" section of 
the manual.

6.  If using Option 2 (4 pin connector Autoswitch) 

check that DIP switch 5 is properly set.  Refer to 
"Configuring the Remote Alarm" section of the 
manual. 

Outputs do not appear to be functioning.

1.  Check the cable to the output device to be sure it 

is properly connected.

2.  Check the Output Connector to be sure it is 

properly plugged into the Alarm.

3.  Check the external output device to make sure it is 

functioning properly.

Serial Port does not work.

1.  Check that you are using the correct cable.  The 

product uses a standard serial cable, not a null 
modem cable.

2.  Check that the PC being used is configured 

properly for the Alarm.

3.  Check that the software used on the PC to 

monitor the Alarm is correct.

SERVICE

Service by a factory authorized facility should be done if an Alarm becomes inoperative and troubleshooting has not 
isolated the problem.  Contact CONCOA Customer Service at 1-800-225-0473 for assistance.

If advised to return the unit for service you should:

1.  Package the Alarm to prevent damage in shipping.  If possible, use the original shipping materials.
2.  Ship the unit back pre-paid.  

3.  Make sure the RMA number appears on the packaging and inside the package.

4.  Include an explanation of the problem you encountered.

Содержание 529 5310

Страница 1: ...talling or Operating Read and Comply with These Instructions Controls Corporation of America 1501 Harpers Road Virginia Beach VA 23454 To Order Call 1 800 225 0473 or 757 422 8330 Fax 757 422 3125 www...

Страница 2: ...s that are brought to the Output Connector so they may be used to control an external device The status indicator lights are organized into pairs representing the Left Bank and Right Bank of an Autosw...

Страница 3: ...mally closed switch One example of such a device would be the CONCOA telephone dialer PN 5295306 Connections to the Alarm are made through input and output connectors Customer supplied cables are easi...

Страница 4: ...AUTOSWITCH To connect the Multi Station Remote Alarm to an Autoswitch system an interface cable must be made The Autoswitch systems are shipped with a connector that is used for this cable An Input C...

Страница 5: ...status RIGHT BANK 5 2 alarm 4 15 RBSW status 3 3 12 6 16 Gnd 2 LEFT BANK 2 4 alarm 1 17 LBSW status RIGHT BANK 5 5 alarm 4 18 RBSW status 3 6 12 6 19 Gnd 3 LEFT BANK 2 7 alarm 1 20 LBSW status RIGHT B...

Страница 6: ...rm the 4 conductor cable already exists Remove the cable from the old alarm by unscrewing the wires from the terminal block Cut the end of the cable so that the wires that went into the terminals on t...

Страница 7: ...BANK 1 12 12 2 24 LBSW status RIGHT BANK 3 23 RBSW status 4 25 Gnd Table 2 CONNECTING THE OUTPUTS The Multi Station Remote Alarm activates relay contacts in response to alarm conditions These relay co...

Страница 8: ...the Alarm will make this information available to the Serial Port as an 8 bit word The data is organized such that each bit represents the status of a BANK Pressure Switch Table 5 shows the bit assig...

Страница 9: ...is limited to the following Command Code Keystroke Definition CAN Hex 18 Cntrl X Cancel ENQ Hex 05 Cntrl E Enquire STX Hex 02 Cntrl B Start of Text ETX Hex 03 Cntrl C End of Text DC4 Hex 14 Cntrl T D...

Страница 10: ...o set the unit on a flat surface With the Alarm placed so that the Input Output connectors are facing towards you unscrew the four 4 screws that hold the cover on They are retained in the cover so the...

Страница 11: ...m the Remote Alarm 2 It is best to set the unit on a flat surface With the Alarm placed so that the Input Output connectors are facing towards you unscrew the four 4 screws that hold the cover on They...

Страница 12: ...ing Option 2 4 pin connector Autoswitch check that DIP switch 5 is properly set Refer to Configuring the Remote Alarm section of the manual Outputs do not appear to be functioning 1 Check the cable to...

Страница 13: ...LECTRICAL SPECIFICATIONS Multi Station Remote Alarm 120 Volt Version 120 Volts 10 50 60 hz 6VA 220 Volt Version 220 Volts 10 50 60 hz 6VA Fuses 120 Volt Version 0 5A normal blow type 3AG 220 Volt Vers...

Страница 14: ...14 THIS PAGE INTENTIONALLY LEFT BLANK...

Страница 15: ...one other than CONTROLS CORPORATION OF AMERICA or if the equipment has been subject to abuse misuse negligence or accident CONTROLS CORPORATION OF AMERICA s sole and exclusive obligation and Buyer s s...

Страница 16: ...Number Description 1 5295310 01 001 120 VAC MULTI STATION REMOTE ALARM 1 5295311 01 001 220 VAC MULTI STATION REMOTE ALARM INCLUDES 1 8308824 120 VAC MULTI STATION REMOTE ALARM 1 8308825 220 VAC MULTI...

Отзывы: