Compur Statox 4120
Procedure:
Remove splash guard and replace it by test gas adapter.
Connect span gas cylinder.
Do not open regulator yet!
Connect diagnostic box to sensor head.
Reading diagnostic box: Red LED flashing; - “ready” – “x.xxx ppm”
Push CAL button Reading diagnostic box: “CALIBR”
Push ENT button Reading diagnostic box: “ppm = ?”
Enter span gas concentration in ppm. Reading diagnostic box: ppm = „value“.
If you have entered a wrong value push button CE and repeat.
If you have entered an invalid concentration, the reading is “ERR KONZ”: Push button CAL
again and repeat.
Now push
button ENT Reading diagnostic box: “GAS ON!”
This is the last opportunity to abort the procedure!
Open span gas regulator
Push button ENT Reading diagnostic box: “WAIT”
Reading of a measured value after 2 minutes. Display diagnostic box: “x.xx ppm” (this not a
calibration value yet).
Wait 1 minute to obtain second measured value from sensor head. Reading diagnostic box:
“x.xx ppm“ (this not a calibration value yet).
Wait 1 minute
This procedure will repeat until the difference between two measured values is within the specified
range. Then the new calibration factor will be displayed for 10 seconds.
Reading diagnostic box: “FKT = x.xx”; “GAS OFF!”
Close span gas regulator
Confirm gas is off by pushing ENT Display diagnostic box: “READY”
Disconnect diagnostic box and switch it off. Close receptacle for opto coupler with the plug. Mount
splash guard.
6.2.2 Use ex – works calibration
Each sensor carries a calibration factor. This factor indicates how far its individual sensitivity differs
from a theoretical value (=1). Values between 0.60 and 2.00 will be accepted by the system. The lower
the factor, the higher its sensitivity. The sensor head microprocessor corrects the sensor signal with
this factor to obtain the correct measured value.
Programming the calibration factor:
Action
Reading Diagnostic box
Switch diagnostic box on
Connect diagnostic box to sensor head
Red LED flashing;
READY-x.xxx ppm
Push FKT
FCT x.xx
Push FKT + ENT (no long break!)
FCT = ?
ERR INPU
Enter factor If you have entered a wrong factor, push CE and
repeat.
FCT = 1.20 (example
FCT = ?
Now push ENT twice immediately
Red LED flashing READY
Switch diagnostic box off, Factor will be transmitted.
Disconnect diagnostic box
24
Содержание STATOX 4120
Страница 1: ...Operation Manual Compur Statox 4120...