Страница 3: ...tions Compliance West USA is proud to present you with your HTT 1S The operating instructions are included in the Manual so please retain this information for future use Thank you for your trust and c...

Страница 4: ...ii...

Страница 5: ...t Panel Features 7 Figure 3 HTT 1S OFFSET switch location 7 TTable 2 HTT 1S front panel Controls Indicators and Connectors 8 Setup Procedure Hipot Tester 9 Setup Procedure Ground Bond Tester 11 Figure...

Страница 6: ...1...

Страница 7: ...at a time to ensure that your testers are functioning correctly A proper test plan would be constituted of the following tests 1 Connect HTT 1S to the tester outputs 2 Set all three switches on the HT...

Страница 8: ...3 problem can be resolved Section 4 of this manual also has troubleshooting information to help you determine if the HTT 1S is functioning correctly...

Страница 9: ...ecifications of the HTT 1S Maximum current capability of GROUND test circuit 30 Amps Ground Test time should be 3 seconds Max and rest 10 s for currents above 10A Setting of PASS for Ground Circuit 0...

Страница 10: ...5 Figure 1 Electrical Diagram of the HTT 1S...

Страница 11: ...HTT 1S Function Tester This manual Return cable Output cable Remove the HTT 1S from the container and place it on a non conductive test bench Plug the cables follow instructions on Set up procedure To...

Страница 12: ...7 Figure 2 HTT 1S Front Panel Features Figure 3 HTT 1S OFFSET switch location...

Страница 13: ...e tester should stop testing and indicate a failure When set to PASS a normal passing result is simulated 5 Leakage Test Switch When set to FAIL an excess leakage condition is simulated and the tester...

Страница 14: ...t fit the hipot ground bond tester For custom cables please contact Compliance West USA 1 Connect the hipot tester to the HTT 1S i On your hipot tester identify the high voltage output receptacle Conn...

Страница 15: ...d GROUND OPEN test by setting the GROUND switch on the HTT 1S to FAIL All other switches on the HTT 1S should be set to PASS Begin testing by starting your hipot tester i Your hipot tester should indi...

Страница 16: ...ound bond tester has an output current above 15A make sure the test time is not higher than 3 seconds to repeat another test wait for 10 seconds Resistors inside may overheat this can cause permanent...

Страница 17: ...IL stop testing and sound a buzzer or otherwise warn the operator of a problem Unanticipated Results The HTT 1S test may cause your tester to give results not consistent with those anticipated by the...

Страница 18: ...easured with a high precision resistance meter Switch designation Switch position Operating mode GROUND LEAKAGE HIPOT PASS PASS PASS No failure the tester connected to the HTT 1S should indicate test...

Страница 19: ...14 Table 4 HTT 1S Resistance for ground circuit Ground switch position Resistance between Chassis and Return with 0 03 tolerances Pass 0 025 0 055 Fail 0 1 0 13...

Страница 20: ...ce West USA is available Phone 800 748 6224 Hours 8 00 AM 4 00 PM Pacific Time Also available on our web site at www compwest com Contact Compliance West USA 650 Gateway Center Way Suite D San Diego C...