background image

 

Section 3 

Operation 

 
This  section  describes  how  to  set  up  and  make  measurements  with  the  HT-10KVDC.    We 
recommend that you read the entire section carefully so that you can use all of its features.

  

 

Setting up the HT-10KVDC 

 
The HT-10KVDC is shipped in a special protective container that should prevent damage during 
shipping.  The container should include the following: 
 

 

The HT-10KVDC Dielectric Withstand Tester 

 

 

A black 18 AWG Test Return Lead (Alligator Clip/Banana Plug ends) 

 

 

A red 18 AWG High Voltage Test Lead (Alligator Clip/High Voltage Plug ends) 

 

 

A Power Cord.  

 

 

This Instruction Manual 

 

Use  the  original  shipping  container  for  subsequent  shipping.    If    the  original  shipping 
container  is  not  available,  be  sure  that  adequate  protection  is  provided  to  prevent  damage 
during shipment. 

 
Remove the Tester from its container and place it on a test bench.   

 

AC Line Voltage Requirements 

 
Connect the HT-10KVDC only to the voltage source per the rating on the rear panel.

 

 

Fuse Replacement 

 

There  is  a  user-replaceable  fuse  (F1)  located  on  the  rear  panel.    The  fuse  rating  is  on  the  rear 
panel. 
 
Use the following procedure to replace the fuse F1: 
1.  Turn the power switch to the O or off position. 
2.  Unplug the HT-10KVDC from the source of supply. 
3.  Remove the power cord from the HT-10KVDC. 
4.  Open the fuseholder door. 
5.  Replace the fuse with one of the correct rating. 
6.  Replace the fuseholder door and power inlet cord. 

Содержание HT-10KVDC

Страница 1: ...HT 10KVDC Dielectric Withstand Tester 0 14000 Volts DC Output Instruction Manual COMPLIANCE WESTUSA...

Страница 2: ......

Страница 3: ...panel and represents the latest in high voltage laboratory testing To fully appreciate all the features of your new meter we suggest that you take a few moments to review this manual Compliance West...

Страница 4: ......

Страница 5: ...justment of the Leakage Current Level 14 Adjustment of the High Voltage Ramp Time 15 Adjustment of the High Voltage Level 15 Adjustment of the High Voltage Test Duration 15 Setting the Voltage Ramp Sw...

Страница 6: ......

Страница 7: ...e supplied cordset Turn it on using the switch on the rear panel 2 Switch the Test Duration switch on the rear panel to the DEFEAT position 3 Push the RESET Button then push and hold the TEST Button T...

Страница 8: ...ost equipment However higher than normal leakage current on a particular sample may indicate an assembly or component problem in the circuit The leakage current is also monitored by the HT 10KVDC to e...

Страница 9: ...d test The HT 10KVDC should remain connected to the circuit until the green Hipot Pass light or the red Hipot Fail light on the front panel is lit and the output voltage as indicated on the front pane...

Страница 10: ...cuit returns output voltage to safe levels before test termination The Hipot Pass or Hipot Fail light will not light until voltage has been ramped down by the HT 10KVDC If a failure is encountered the...

Страница 11: ...p ON DEFEAT Timer Duration ON DEFEAT Failure Shutdown ON DEFEAT ENVIRONMENTAL Operating Temperature 15 40 C Relative Humidity Range 0 90 non condensing GENERAL Input power requirements Model HT 10KVDC...

Страница 12: ...iginal shipping container for subsequent shipping If the original shipping container is not available be sure that adequate protection is provided to prevent damage during shipment Remove the Tester f...

Страница 13: ...and Rear Panel Features The front panel features of the HT 10KVDC are shown in Figure 3 1 and described in Table 3 1 The rear panel features of the HT 10KVDC are shown in Figure 3 2 and described in T...

Страница 14: ...9 Figure 3 1 Controls Indicators Connectors HT 10KVDC Front Panel...

Страница 15: ...l light and if not defeated the high voltage duration time starts when the voltage output reaches the preset level 5 Hipot Fail LED Indicates failure of high voltage test If arcing or a flashover of t...

Страница 16: ...11 Figure 3 2 Controls Indicators Connectors HT 10KVDC Rear Panel...

Страница 17: ...e test duration See Adjustment of the High Voltage Test Duration section 16 Leakage Limit Adjustment Adjusts the alarm level for the excessive leakage current test See Adjustment of the Leakage Curren...

Страница 18: ...mps should be lit 10 Disconnect the black lead from the red lead and remove both leads from the HT 10KVDC 11 Enable the voltage output by pressing the RESET button then the TEST button After the full...

Страница 19: ...st the Leakage Limit control item 16 on the rear panel to its maximum clockwise position 4 Make sure there are no test leads connected to the Tester 5 Push the RESET button item 1 The blue TEST light...

Страница 20: ...n the Test Timer switch item 12 to the DEFEAT position The HT 10KVDC will supply voltage while the TEST button item 2 is held in 2 Make sure there are no test leads connected to the HT 10KVDC 3 Press...

Страница 21: ...testing after a failure is encountered This allows the operator to find a breakdown point but all safety shutdown circuitry in the HT 10KVDC is disabled when the Failure Shutdown switch is in the DEFE...

Страница 22: ...t of the circuit under test is too high the Excess Leakage LED item 3 will light and the test will terminate If the Failure Shutdown Switch is set to DEFEAT and the requirements of Table 3 3 are met t...

Страница 23: ...er test result means the equipment being tested failed a test phase If unanticipated test failures continue and you suspect that the equipment under test is built correctly check the following item 1...

Страница 24: ...19 Voltage Ramp ON DEFEAT ON DEFEAT DEFEAT Test Timer ON ON DEFEAT DEFEAT DEFEAT Failure Shutdown ON ON ON ON DEFEAT See Below A B C D E Table 3 3 Rear Panel Switch Truth Table...

Страница 25: ...will stop automatically on all leakage or breakdown failures For safety we recommend that you begin testing with the front panel voltage knob set at minimum E Total Defeat Full voltage is produced at...

Страница 26: ...00 748 6224 Technical Assistance is available from Compliance West USA between the hours of 8 30 AM and 4 30 PM Pacific Time Compliance West USA 2120 Jimmy Durante Blvd Suite 118 Del Mar CA 92014 Phon...

Страница 27: ...rvice Information The HT 10KVDC is warranted to the original purchaser for a period of 1 year This warranty does not cover problems due to misuse or neglect Malfunctions which occur within the limits...

Страница 28: ...8 and 9 4 Push the red RESET button item 1 The blue TEST button will light item 2 5 Push the blue TEST button 6 The Tester will conduct a test sequence The meter will show the output voltage hold and...

Страница 29: ...bilize for approximately five minutes Perform all calibration adjustments at an ambient temperature of 23 C 5 C 73 F 9 F WARNING CALIBRATION ADJUSTMENTS ARE PERFORMED ON ENERGIZED CIRCUITS EXERCISE CA...