Compliance Products USA, Inc
CAL-Fac-001-MP-CDT 240 Rev 2.3
Page 8 of 13
6.2.2.
Edit the test parameters for a Voltage test:
•
Trigger on NEG
•
Voltage Test
•
Voltage setting: 50.0V
•
Neutral to Ground measurement
6.2.3.
Press TEST once to ARM the CDT.
6.2.4.
Record both input voltage and N-G BNC voltage on section 4 of CAL-Log-001-CDT 240.
6.2.4.1.
EUT Test Voltage: DMM2 Neutral to Ground voltage readout.
6.2.4.2.
Expected Value:
_
_
_
100
EUT Test Voltage
Expected
value
=
6.2.4.3.
Tolerance: Expected Value +/-1 % of reading.
6.3.
Line to Neutral - No adjustment, verification only.
6.3.1.
If the EUT connection is reverse, connect as normal
6.3.2.
Edit the test parameters for a Voltage test:
•
Trigger on NEG
•
Voltage Test
•
Voltage setting: 50.0 V
•
Line to Neutral measurement
6.3.3.
Press TEST once to ARM the CDT.
6.3.4.
Record both input voltage and L-N BNC voltage on section 4 of CAL-Log-001-CDT 240.
6.3.4.1.
EUT Test Voltage: DMM2 Line to Neutral voltage readout.
6.3.4.2.
Expected Value:
_
_
_
100
EUT Test Voltage
Expected
value
=
6.3.4.3.
Tolerance: Expected Value +/-1 % of reading.