Emitted electromagnetic interference
The Comper Smarkin is intended for use in the electromagnetic
environment specified below. The customer or the user of the Comper
Smarkin should confirm that it is used in such an environment..
Table 1: Electromagnetic Environment specified
Emissions test
Compliance
Electromagnetic environment
guidance
RF emissions CISPR 11
Group 1, Class A
The Comper Smarkin uses RF energy
only for its internal function. Therefore,
its RF emissions are very low and are not
likely to cause any interference in nearby
electronic equipment.
Harmonic emissions IEC
61000-3-2
Class A
Voltage fluctuations/ flicker
emissions
IEC 61000-3-3
Complies
Electromagnetic immunity
.
Emissions test
Compliance
Electromagnetic environment
guidance
Electrostatic discharge (ESD)
IEC 61000-4-2
6kV contact 8kV air
±
2kV,
±
4kV,
±
6kV
contact
±
2kV,
±
4kV,
±
8kV air
discharge
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
Electrical fast transient/
burst IEC 61000-4-4
±
2kV for power supply lines
±
2kV for power
supply lines
Mains power quality should be that
of a typical commercial or hospital
environment.
Surge
IEC 61000-4-5
±
1kV Line to line
±
2kV Line to earth
±
1kV Line to
line
±
2kV Line to
earth
Mains power quality should be that
of a typical commercial or hospital
environment.
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
IEC 61000-4-11
<5% UT(>95%
dip
in UT)
for 0,5 cycle
40% UT(60% dip
in UT)
for 5 cycles 70%
UT (30% dip in
UT) for 25 cycles
<5% UT
(>95% dip in UT)
for 5 sec
Mains power quality should be that
of a typical commercial or hospital
environment. If the user of the Comper
Smarkin requires continued operation
during power mains interruptions, it is
recommended that the Comper Smarkin
be powered from an uninterruptible
power supply or a battery.
Power frequency (50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
Power frequency magnetic fields should
be at levels characteristic of a typical
location in a typical commercial or
hospital environment.